5-10
Maintenance Procedures
Fitting the source enclosure to the instrument
Required materials
Chemical-resistant, powder-free gloves
To fit the source enclosure to the instrument:
1. Using both hands, fit the source enclosure to the two supporting studs
on the source adaptor housing.
2. Close the source enclosure.
3. Connect the probe adjuster and options cables to the instrument’s
connectors.
Tip:
The cables and connectors are color coded; the blue-sleeved cable
connects to the blue connector and the yellow-sleeved cable to the yellow
connector.
4. Slide closed the instrument’s source interface door.
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
To avoid puncture wounds, take great care while fitting the
source enclosure to the source if a corona pin is fitted (the pin tip is
sharp).
Caution:
If you are fitting a NanoLockSpray source enclosure, to
avoid damage to the sample inlet, you must move the sprayer
platform out of the source enclosure before you close the source
enclosure (see
).
Содержание Xevo G2 QTof
Страница 10: ...x...
Страница 18: ...xviii Table of Contents...
Страница 34: ...1 16 Waters Xevo G2 QTof Overview...
Страница 44: ...2 10 Preparing the Mass Spectrometer for Operation...
Страница 108: ...5 32 Maintenance Procedures 2 Slide the collar to the end of the tool 3 Insert the collar in the sample cone Collar...
Страница 227: ...Connecting the liquid waste line B 17 Positioning of drain tube TP01807 Correct Incorrect...
Страница 238: ...B 28 External Connections...
Страница 242: ...C 4 Materials of Construction and Compatible Solvents...
Страница 248: ...Index 6...