Configuring the NanoLockSpray Source
Advancing and retracting the sprayer platform
To move the sprayer platform out of the source:
1. Confirm that the sprayer’s safety cover is installed (see the figure on
).
2. Unscrew the thumbscrew on the front of the sprayer platform.
3. Pull out the side thumbscrew and withdraw the sprayer platform from
the source.
4. Release the side thumbscrew, locking the platform in the withdrawn
position.
To move the sprayer platform into the source:
1. Confirm that the sprayer’s safety cover is installed (see the figure on
).
2. Pull out the side thumbscrew, and push the sprayer platform into the
source.
3. Release the side thumbscrew, locking the platform in position.
4. Tighten the front thumbscrew, securing the adjuster assembly rigidly to
the source.
Warning:
To avoid electrical shock, ensure the safety cover is in place
over the sprayer.
Warning:
To avoid electrical shock, ensure the safety cover is in place
over the sprayer.