Maintenance schedule
5-3
Maintenance schedule
The following table lists periodic maintenance schedules that ensure optimum
instrument performance.
Maintenance schedule:
Procedure
Frequency
For information...
Clean the instrument case.
As required.
Empty the nitrogen exhaust
trap bottle.
Check daily, empty as
required.
Inspect and adjust the
roughing pump’s oil level.
Weekly.
Replace the roughing pump’s
oil and oil mist filter.
Annually.
See
Replace the oil-free (scroll)
pump’s seals.
Annually.
See Edwards
document
XDS35i
Instruction Manual
A730-01-880
.
Clean the source components. When sensitivity
decreases to
unacceptable levels.
Replace the ESI probe tip.
When sensitivity
decreases to
unacceptable levels.
Replace the ESI probe
capillary.
When sensitivity
decreases to
unacceptable levels or
sample flow is
inconsistent.
Clean the APCI probe tip.
(Options using the APCI
probe only.)
When sensitivity
decreases to
unacceptable levels.
Replace the APCI probe
capillary.
When sensitivity
decreases to
unacceptable levels or
sample flow is
inconsistent.
Содержание Xevo G2 QTof
Страница 10: ...x...
Страница 18: ...xviii Table of Contents...
Страница 34: ...1 16 Waters Xevo G2 QTof Overview...
Страница 44: ...2 10 Preparing the Mass Spectrometer for Operation...
Страница 108: ...5 32 Maintenance Procedures 2 Slide the collar to the end of the tool 3 Insert the collar in the sample cone Collar...
Страница 227: ...Connecting the liquid waste line B 17 Positioning of drain tube TP01807 Correct Incorrect...
Страница 238: ...B 28 External Connections...
Страница 242: ...C 4 Materials of Construction and Compatible Solvents...
Страница 248: ...Index 6...