5-74
Maintenance Procedures
20. Replace the combined 2.5-mm Allen wrench and cone extraction tool in
its storage location on the source adaptor housing.
21. Fit the probe heater (see
).
22. Fit the probe to the instrument (see
).
23. In the Instrument Console, click API
to start the probe and
desolvation gas flows.
Replacing the LockSpray probe capillary
Replace the LockSpray probe capillary if it becomes blocked and you cannot
clear it, or if it becomes contaminated or damaged.
Removing the existing capillary
Required materials
•
Chemical-resistant, powder-free gloves
•
4-mm wrench
•
7-mm wrench
Caution:
• When handling the probe heater, take great care to grip the
heater so as not to damage its electrical wiring.
• Take great care not to damage the probe heater’s electrical
connections, capillary sleeve, or capillary when fitting the
heater over the capillary sleeve.