3-14
Configuring the LockSpray Source
25. Replace the combined 2.5-mm Allen wrench and cone extraction tool in
its storage location on the source adaptor housing.
26. Fit the ESI probe to the source (see
).
Configuring for APCI mode
To operate in APCI mode, you must fit the APCI probe and corona pin to the
LockSpray source enclosure.
For more information on using APCI mode, see the SYNAPT G2 MS system
online Help.
Installing the APCI probe
Required materials
•
Chemical-resistant, powder-free gloves
•
PEEK tubing
To install the APCI probe
Warning:
The LC system connections, APCI probe, and source can
be contaminated with biohazardous and/or toxic materials. Always
wear chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
To avoid electric shock, prepare the instrument for work
performed on its source before commencing this procedure.