Waters CORTECS Скачать руководство пользователя страница 10

[ CARE AND USE MANUAL ]

Waters Corporation 

34 Maple Street 
Milford, MA 01757 U.S.A. 
T: 1 508 478 2000 
F: 1 508 872 1990 
www.waters.com

[ CARE AND USE MANUAL ]

Waters, The Science of What’s Possible, CORTECS, ACQUITY, UPLC, eCord, VanGuard, MaxPeak, Arc, 
Oasis, Sep-Pak, Empower, and MassLynx are trademarks of Waters Corporation. 
All other trademarks are the property of their respective owners.

2022 Waters Corporation. Produced in the U.S.A.  December 2022  720005024EN  KK-PDF

Injection Solvents

1.  If possible, injection solvents should be 95% acetonitrile. 

The polar solvent (i.e., water, methanol, isopropanol) 
should be minimized to 25% of the total volume.

2.  A generic injection solvent is 75:25 acetonitrile:methanol. 

This is a good compromise between analyte solubility and 
peak shape.

3.  Avoid water and dimethylsulfoxide (DMSO) as injection 

solvents. These solvents will produce very poor peak shapes.

4.  Exchange water or DMSO with acetonitrile by using 

reversed-phase solid-phase extraction (SPE). If this is not 
possible, dilute the water or DMSO with organic solvent.

Miscellaneous Tips

1.  CORTECS HILIC columns are designed to retain very polar 

bases. Acidic, neutral, and/or non-polar compounds 
will have limited retention.

2.  Optimal flow rates for small (<200 daltons) very polar 

bases are in the 0.4 to 0.8 mL/min range.

3.  As compared to Atlantis™ HILIC Silica HPLC Columns, 

the CORTECS HILIC columns are approximately 20% 
less retentive for gradient analysis and 35–65% less 
retentivefor isocratic analysis.

4.  In HILIC, it is important to remember that water is the 

strongest solvent. Therefore, it must be eliminated or 
minimized in the injection solvent.

5.  For initial scouting conditions, run a gradient from 95% 

acetonitrile to 50% acetonitrile. If no retention occurs, 
run isocratically with 95:3:2 acetonitrile:methanol: 
aqueous buffer. 

6.  Alternate polar solvents such as methanol, acetone, 

or isopropanol can also be used in place of water to 

increase retention.

7.  If using an ACQUITY UPLC System, the weak needle wash 

should closely match the % organic present in the initial 
mobile-phase conditions, otherwise, analyte peak shape 
or retention may suffer.

8.  It is recommended to monitor your systems health by 

incorporating the HILIC Quality Control (QC) Reference 
Material into your work flow. (Reference 

720004535EN

).

VII. CAUTIONARY NOTE 

Depending on the user’s application, these products may be 
classified as hazardous following their use, and as such are 
intended to be used by professional laboratory personnel 
trained in the competent handling of such materials. 
Responsibility for the safe use and disposal of products rests 
entirely with the purchaser and user. The Safety Data Sheet 
(SDS) for this product is available at 

www.waters.com/sds

Содержание CORTECS

Страница 1: ...CORTECS packing material is tested chromatographically with acidic basic and neutral analytes and the results are held to narrow specification ranges to assure excellent reproducible performance Ever...

Страница 2: ...lumn outlet to the detector This prevents entry of air into the detection system 4 Gradually increase the flow rate as described in Step 2 5 Once a steady backpressure and baseline have been achieved...

Страница 3: ...guard holder and guard cartridge can be easily attached to the inlet of the column For more information on VanGuard products visit www waters com vanguard g VanGuard FIT Cartridges The VanGuard FIT C...

Страница 4: ...ating temperatures are recommended for longer column lifetime Working at high temperatures e g 45 C may also result in shorter column lifetimes III COLUMN USE To ensure the continued high performance...

Страница 5: ...at 10 mM or less Ammonia NH4 OH Ammonium bicarbonate 9 2 10 3 HCO3 9 2 NH4 8 2 10 2 8 2 11 3 Volatile Volatile Yes Yes Used in the 5 10 mM range for MS work keep source 150 C Adjust pH with ammonium...

Страница 6: ...rface charge modification End cap style Carbon load Ligand density pH limits Temp limits C18 Trifunctional C18 None Proprietary 6 6 2 7 mol m2 2 8 45 C C18 Trifunctional C18 Proprietary 5 7 2 4 mol m2...

Страница 7: ...rmate containing mobile phase V eCORD INTELLIGENT CHIP TECHNOLOGY May not be available for all column configurations a Introduction The eCord Intelligent Chip provides the history of a column s perfor...

Страница 8: ...ple set maximum pressure and temperature in the sample set and if the column met basic system suitability requirements Up to 50 sample sets can be stored on the eCord chip VI ADDITIONAL INFORMATION a...

Страница 9: ...related questions to ask If I inject neat standards prepared in mobile phase do I observe these problems If I prepare my standards in water and prepare them like samples e g SPE filtration etc do I s...

Страница 10: ...es are in the 0 4 to 0 8 mL min range 3 As compared to Atlantis HILIC Silica HPLC Columns the CORTECS HILIC columns are approximately 20 less retentive for gradient analysis and 35 65 less retentivefo...

Отзывы: