To replace a solvent reservoir filter:
1.
Remove the filtered end of the solvent tubing from the solvent bottle.
2.
Remove the old solvent reservoir filter from the short piece of fluoropolymer tubing.
3.
Insert the new solvent reservoir filter into the fluoropolymer tubing, pushing until it contacts
the solvent tubing.
Figure 3–2: Replacing the solvent reservoir filter
1
2
3
Solvent tubing
Fluoropolymer tubing
Solvent reservoir filter
4.
Insert the filtered end of the solvent tubing into the solvent bottle.
5.
Shake the solvent tubing to remove any air from the filter.
6.
Prime the solvent manager (see
3.9 Replacing the leak sensor
Warning:
Observe Good Laboratory Practice (GLP) at all times, particularly when
working with hazardous materials. Consult the Material Safety Data Sheets regarding
the solvents you use. Additionally, consult the safety representative for your
organization regarding its protocols for handling such materials.
Warning:
To avoid personal contamination with biohazards or compounds that are toxic, wear
clean, chemical-resistant, powder-free gloves when performing this procedure.
Requirement:
Wear clean, chemical-resistant, powder-free gloves when performing
this procedure.
August 8, 2016, 715005050 Rev. C
Page 48