Required material:
•
Chemical-resistant, powder-free gloves
•
Bottle-stopper assembly
•
2.5-L bottle
•
Cable tie
To connect to the 2.5-L waste container:
1.
Connect the bottle-stopper tube assembly’s 4-mm outer-diameter (O.D.) drain tubing to the
4-mm bulkhead fitting on the instrument’s rear panel (see the figure below).
Tip:
To ensure that excess tubing remains organized when the instrument is positioned at
the bottom of an LC stack, loop the tubing, taking care not to kink it, and secure it using a
cable tie. Alternatively, if the instrument is unlikely to change location, cut the tubing to
length.
2.
If you are using a Standard QDa Detector, connect the 6-mm O.D. tube from the
bottle-stopper assembly to the diaphragm pump’s 6-mm elbow (as shown in the following
figure).
Figure D–13:
Waste connections
Requirement:
Wear clean, chemical-resistant, powder-free gloves when connecting the
waste container.
6-mm O.D. tubing
(Standard QDa only)
6-mm elbow
Diaphragm pump
(Standard QDa only)
ACQUITY QDa
Detector
4-mm bulkhead fitting
(KAB only)
4-mm O.D. drain tubing
(KAB only)
Route both tubes to the bottle-stopper assembly
on the 2.5-L waste container