715001213, Rev. B
3 of 5
A. Remove the nebulizer and fill the U-tube with mobile phase.
B. Slide the nebulizer into place with the air coupler at 12 o’clock to align the pins
and turn it 90 degrees clockwise.
C. Connect the column’s outlet using PEEK tubing (P/N 430001562) to the
nebulizer’s inlet port. If connecting from a UV detector, use PEEK tubing P/N
430001565.
D. Connect the gas input to the nebulizer’s gas inlet port.
E. Connect the drip tray tubing to the waste management system.
E
F
B
C
D
Push in, then
turn right 90
degrees.
Install tubing.
Nebulizer Connections
A
Figure 2 - Filling U-Tube
Figure 3 - Nebulizer Connections
Figure 4 - Plumbing and Leak Sensor Installed