Connecting power and signal cables
2-5
Required materials
•
If using Ethernet communications, an Ethernet cable, and optionally, an
Ethernet switch (supplied with the Ethernet Switch Communications
Kit, part number 700004123).
•
If using IEEE-488 communications, a 2-meter IEEE-488 cable (supplied
with the data control system).
•
Power cord (startup kit)
•
Flat-blade screwdriver – small (required when connecting the inject
start signal cable)
Connecting the power supply
The 1500-series pump automatically adjusts for AC input voltage.
To make the power supply connection:
1. Insert the 120 V or 240 V power cord into the power connector on the
rear of the pump.
2. Plug the other end of the power cord into a grounded power outlet.
Making ethernet connections
The 1500-series pump, other chromatography equipment in the HPLC system
such as the column heater, autosampler, detector, and the data control
workstation can interconnect using Ethernet cables and a networking switch
).
The pump is equipped with a RJ-45 connector for Ethernet port
communications. The Ethernet port—a 10/100 Base-T networking interface, is
used only for remote control, the case when Empower 2 controls its operation,
and for firmware upgrades via the Waters Autoloader utility.
Requirements:
•
In an Ethernet configuration, all Waters HPLC system components,
including the 1500-series pump and optional column heater, must
communicate with the data system via Ethernet communications.
•
As with IEEE-488 control, when using an autoinjector, triggering of the
inject start signal for the 1500-series pump occurs over the Ethernet
Содержание 1500-Series
Страница 14: ...xiv Table of Contents ...
Страница 32: ...1 18 Introduction ...
Страница 50: ...2 18 Installing the HPLC Pump ...
Страница 84: ...3 34 Installing Options and Accessories ...
Страница 108: ...4 24 Preparing for Operation ...
Страница 160: ...A 14 Safety Advisories ...
Страница 184: ...C 16 Solvent Considerations ...