background image

 

WL800

 Manual 

                                                                         Page 63 – Revision 12‐11‐2018 

 

FAULT CODE INDEX 

 

1.  UV LED Blinking – UV Error 

 

2.  Heating and Service LED Blink Simultaneously – Hot Water Sensor Error 

 

3.  Chilling and Service LED Light Blink Simultaneously – Cold Water Sensor Error 

 

4.  Heating and Service LED Alternate Blinking –Low Water Level Error 

 

5.  Chilling and Service LED Alternate Blinking – Full Water Level Error 

 

6.  Heating, Chilling and Service LED Sequentially Blink 

 

7.  Service LED is On – Leak Detection Probe Has Detected Water 

  

 
UV LED Blinking – UV Error 
 

Possible Reason 

Solution 

UV Lamp or Pump is Not 
Operational 

Check Connections to UV Lamp / Pump Assembly.  Replace 
UV Lamp / Pump Assembly as needed. 

 
 
Heating and Service LED Blink Simultaneously – Hot Water Sensor Error 

 

Possible Reason 

Solution 

Sensor not connected or is not 
operating within set 
parameters 

1.

 

Check connection to Hot water sensor. 
 

2.

 

Check OHMS reading – above 900K indicates open 
sensor.  Below 5.2K indicates sensor short. 

 
Replace Hot Water Sensor. 

 
 
Chilling and Service LED Light Blink Simultaneously – Cold Water Sensor Error 

 

Possible Reason 

Solution 

Sensor is not connected or is 
not operating within set 
parameters 

1.

 

Check connection to Cold Water Sensor. 

2.

 

Check OHMS reading – above 690K Indicates Open 
Sensor, below 7.3K indicates Sensor Short. 

 

Replace Cold Water Sensor. 

 

Отзывы: