Calibration:

< pH >

1.  Dip the electrode into the buffer solution pH 7. 
 

Stir gently and wait until the display stabilized. 

 

Adjust the reading to 7.0 (6011 only) or 

 

7.00 (6012 only) at 25 

 by turning the 

 

trimmer (ZERO) located at right side of battery 

 

compartment with a screwdriver.

2.  Rinse the electrode with clean water and wipe 
 

it dry. Dip the electrode into the buffer solution  

 

pH4.Stir gently and wait until the display 

 

stabilized. Adjust the reading to 4.0 (6011 

 

only) or 4.00 (6012 only) at 25 

 by turning 

 

the trimmer (Span) located at left side of 

 

battery compartment with a screwdriver.

5

6

Accessories:

Upon receiving the shipment, inspect the 
container and equipment for any signs of damage. 
Remove the packing list and verify that you have 
received all equipments:
Meter, Buffer pH 4 & 7 & Soaking solution, Battery 
(has been installed), Instruction manual, Gift box.

Preparation:

1.  Remove the protection cap and unscrew 
 

soaking bottle from meter to rinse the 

 

electrode with clean water and wipe it dry. 

 

Don't leak soaking solution from bottle, and 

 

replace bottle when end of usage.

2.  Open battery compartment cap to take out the 
 

screwdriver. 

3.  Press POWER button to turn the meter on.