background image

SRM200 User's Manual 

 

 

Part No.: 7100.SRM200

 

 

   

   

Page 4 / 8 

 

V0521 

 

LCD display overview 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

n

 

Measuring Surface Resistance 

·

  To measure the surface resistance of an object, hold the instrument onto the surface and press the 

"PUSH TO TEST" button. 

·

  The value is indicated on the display. The coloured LED's additionally indicate the measuring range if 

no limit is assigned. If the limit is assigned, the limit arrows on the display will indicate the measured 
value being below or above the limit range.    

·

 

▼▲

 selects the storage folder; 

OK

 stores the current measurement value to the selected folder. 

 
 
 
 
 
 
 
 
 
 
 

n

 

Measuring Resistance to Ground 

·

  Plug in the supplied grounding cord at one socket of the instrument. The associated internal electrode 

will be disconnected.  

·

  Connect the opposite end of the grounding cord to "ground" or a "groundable point".  

·

  Hold the instrument onto the surface like described above and press the button. 

 
 
 

 
 
 
 
 
 
 

Measured value 

Relative humidity 

Battery status 

Date 

Temperature

(

V

) View result mode active 

(

D

) Delete mode active

Value exceeds  
selected limit 

Value is below 

selected limit

Assigned limit 

Time 

Storage folder 

PUSH TO TEST 

PUSH TO TEST 

Содержание SRM 200

Страница 1: ...Wolfgang Warmbier GmbH Co KG Systeme gegen Elektrostatik Untere Gießwiesen 21 78247 Hilzingen Germany www warmbier com User s Manual Surface Resistance Meter SRM 200 Firmware 2 x ...

Страница 2: ...e 4 n Measuring Resistance to Ground 4 n Other Measurements 5 TEST VALUES STORAGE 5 FUNCTIONS 5 n View results 5 n Delete results 5 n Change Limits 5 n Folder names 6 n Timeout 6 n Temperature 6 n Date 6 n Calibration 6 n Language 6 PACKING LIST 7 TECHNICAL DATA 7 SPARE PARTS 7 TROUBLE SHOOTING 7 NOTICE 8 MAINTENANCE 8 WARRANTY 8 REPAIR 8 CALIBRATION 8 WASTE DISPOSAL 8 ...

Страница 3: ...W 9x1010 W Electrostatic dissipative Red 1x1011 W Electrostatic insulating 4 Buttons Up Down MENU OK Button Function MENU Open menu Return from sub menu OK Confirm or change value Increase value Scroll down in menu Decrease value Scroll up in menu Simultaneously pressing turns the instrument off 5 Push to test button to switch ON and start measurement 6 USB connector for battery charger and PC con...

Страница 4: ...bove the limit range selects the storage folder OK stores the current measurement value to the selected folder n Measuring Resistance to Ground Plug in the supplied grounding cord at one socket of the instrument The associated internal electrode will be disconnected Connect the opposite end of the grounding cord to ground or a groundable point Hold the instrument onto the surface like described ab...

Страница 5: ...Adjust Date and time Functions The following functions are available Most of them can be accessed more conveniently by the PC software n View results MENU press button View results select OK confirm select folder 1 99 OK confirm folder select record 1 99 OK display value n Delete results MENU press button Delete results select OK confirm select folder 1 99 OK confirm folder select record 1 99 OK c...

Страница 6: ...t OK change value MENU back to menu n Temperature MENU press button Temperature select temperature OK change between C or F MENU Z back to menu n Date MENU press button Date select date OK change date increase decrease value OK accept value MENU back to menu n Calibration Calibration date and firmware version display MENU press button Calibration select calibration OK display MENU back to menu n L...

Страница 7: ...0 x 35 mm L x B x H PC interface USB 2 0 Case ABS Weight 290 g Measuring range Display range Resolution Accuracy 103 W 1x103 9x103 1 kW 10 reading 104 W 1x104 9x104 10 kW 10 reading 105 W 1x105 9x105 100 kW 10 reading 106 W 1x106 9x106 1 MW 10 reading 107 W 1x107 9x107 10 MW 10 reading 108 W 1x108 9x108 100 MW 10 reading 109 W 1x109 9x109 1 GW 10 reading 1010 W 1x1010 9x1010 10 GW 25 reading 1011 ...

Страница 8: ...Replace only a rechargeable battery of the same type and take care of the polarity Warranty The warranty does not include the rechargeable battery battery damage due to drainage and mechanical damage of the instrument The warranty is void if the unit is opened Repair Repairs shall be carried out by qualified personnel only In case you send the instrument for repair please pack it safely and state ...

Отзывы: