n

 

Signals and possible error conditions 

 

 

Green LED flashes 

  Resistance and capacity are in the rated range

 

 

 

 

Red LED flashes  

Reason:  The measured resistance is too high 

Measures:  Check if the wrist strap is fixed tight to the skin 

  Check if grounding wire or the coil cord are broken 

 

 

 

Red LED is permanent on 

Reason 1:  The measured resistance is too low 

Measures 1:  Check if the coil cord contains a safety resistor 

Check for an insulation failure or a to big capacity to conductive table mats 

Reason 2:  Direct  connection  between  terminal  1  and  2  is  missing,  connection  to 

conductive fixtures is broken or one of the earth wires is broken 

Measures 2:  Check connections to terminal 1 and 2 

 

n

 

Calibration 

 
A Calibration Unit is available to check the Safety pips 
correct function and values (Part no.: 7100.181.C) 
 
 
 
 
 
 
 
 
 