23
Do not place the handpiece with cable and the instrument changer in liquid disinfectant or in an ultrasonic
bath.
> Clean the handpiece with cable, the instruments and the instrument changer under running tap water
(< 35°C / < 95°F).
> Rinse and brush off all internal and external surfaces.
> Remove liquid residues using compressed air.
Instruments
Clean and disinfect diamond coated instruments in an ultrasonic bath.
Evidence of the medical device’s basic suitability for effective manual cleaning and disinfection was provided
by an independent test laboratory using the “Bandelin Type RK 100CC” ultrasonic bath and the cleaning agent
and disinfectant “StammopurDR8” (DR H Stamm, Berlin) and “CaviCide™” (Firma Metrex).