Hygiene and maintenance
Manual cleaning
Measuring probe with cable / Universal support
Do not immerse the measuring probe with cable and the universal support in liquid disinfectant or in an ultrasonic bath.
Measuring probe with cable / Universal support
> Clean the measuring probe with cable and the universal support under running tap water (< 35°C / < 95°F).
> Rinse and brush off all internal and external surfaces.
> Remove liquid residues using compressed air.
Osstell ISQ module
> Do not immerse the Osstell ISQ module in water or clean them under running water.