LISA | Instructions for Use | VA131 2019 ENG Rev06 | 16/04/2020 | © 2020 W&H Sterilization Srl
69
Sterilization cycles
4
Check the message. See
"Messages of a stopped
sterilization cycle" below.
5
Tap
to view the cycle
parameters. See "View the cycle
parameters" on page 66.
6
Open the chamber door.
7
Reprocess the load if necessary.
CAUTION! Hot steam. Wait
the steam to dissipate
before opening the door.
Note: Water could be present in the
chamber when opening the door. To
prevent spilling place a towel below the
chamber door.
MESSAGES OF A STOPPED STERILIZATION CYCLE
Following are the messages:
n
Load not sterile: Do not use items on patients!
n
Drying interrupted: The load might be wet. Wet items are for
immediate use only!
Unloading
WARNINGS
CAUTION! Risk of burns. Before touching, ensure the
sterilization chamber is cold. Always use the tray holder.
Sterilization cycle report
WHERE CYCLE DATA ARE STORED
The sterilizer stores in memory the summarized reports of the last
400 cycles and the analytical reports of the last 50 cycles. All
reports can also be saved on the USB pen drive or in a specific
remote folder in the network if the sterilizer is connected to a LAN .
STORED REPORT FORMAT
The summarized reports are stored in HTML format and the
analytical reports in SCL format. All parameters are recorded every
second.