9. Error messages
Icon
Description of error
Solution
WARNING FOOT CONTROL
> Check plug contacts of foot control
> Check the plug contacts of the dongle
WARNING MOTOR
> Check the plug contacts of the motor
> Allow motor to cool for at least 10 minutes
WARNING STORAGE DEVICE
> Insufficient memory available
> Unknown file system
> The write protection is active
> Unknown storage device
> Plug in a USB stick with sufficient memory
WARNING OVERHEATING
> Switch off the control unit
> Allow the control unit to cool for at least 10 minutes
> Switch on the control unit
The error message disappears when it is clocked or when the pedal (grey) on the foot control is released.