Lubrication
> Lubricate the dry medical device immediately after cleaning and/or disinfection.
Recommended lubrication cycles
> Essential after every internal cleaning
> Before each sterilization
or
> After 30 minutes of use or at least once daily
> Chucking system once a week
With W&H Service Oil F1, MD-400
> Follow the instructions on the oil spray can and on the packaging.
or
With W&H Assistina
> Follow the instructions in the Assistina Instructions for use.