37
Proximity fields from RF wireless communications equipment
IEC/EN 60601-1-2 Table 9
IEC/EN 61000-4-3
385 MHz
27 V/m
450 MHz
28 V/m
710 / 745 / 780 MHz
9 V/m
810 / 870 / 930 MHz
28 V/m
1720 / 1845 / 1970 / 2450 MHz
28 V/m
5240 / 5500 / 5785 MHz
9 V/m
Electrical fast transient/burst IEC/EN 61000-4-4
Electrical cables
Input and output cables
±2 kV
±1 kV
Surges
IEC/EN 61000-4-5
–
Conducted disturbances induced by RF fields
IEC/ EN 61000-4-6
3 V
6 V in ISM bands 6 V in amateur radio bands
Power frequency magnetic field
EN 61000-4-8
30 A/m
Voltage dips, voltage interruptions
IEC/EN 61000-4-11
–
Содержание EM-19