Maintenance
68
Lara | Instructions for Use | RIL-1 W&H Med ENG Rev06 | 14/04/2022 | © W&H Sterilization Srl
400-cycle maintenance
REPLACING THE BACTERIOLOGICAL FILTER
Notice
: If you replace this consumable before the maintenance due date you have to reset the cycle counter. See "Replace the consumable
before the maintenance due date" on page 62.
Proceed as follows:
1
Open the service door.
2
Unscrew the bacteriological filter by hand (counter-clockwise).
3
Screw on the new bacteriological filter (clockwise) and tighten it snug.