Chapter 7
Touchscreen Operation
72
Meas. Light Sett.
7.3.1.2
This command opens the menu
“Measuring Light Settings” in which
measuring light intensity (Meas.
Light Int.) can be marked by the
Up/Down arrows, selected by
the SET key and then adjusted us-
ing the arrow keys again. At con-
stant frequency, measuring light in-
tensity can be considered as pro-
portional to intensity settings 1 to 12.
The two remaining items of the menu “Measuring Light Settings”
concern measuring light frequency. Selecting the line “Meas.
Light. Freq.” by SET allows choosing between 5 frequency lev-
els using arrow keys. Table 8 (page 72) lists the measuring light
frequencies of settings 1 to 5.
Table 8: Measuring Light Fre-
quencies
Setting
Frequency, Hz
1
5
2
10
3
15
4
20
5
25
high
100
PAM Settings
A
OFF
SET
EXIT
Side Panel
ON/OFF
. Switch for
measuring light.
,
. Up/down
keys move curser.
SET
. Open submenu
(
) or select parameter,
change parameter value
by up/down keys.
EXIT
. Return to Main
Menu.
Meas. Light
On
Meas. Light Sett.
Gain
1
Damping
2
ETR-Factor
0.84
Fo’ Mode
On
Adjust F-Offset
Ft
421
Current F-Offset
70
Fig. 39:
PAM Settings