Touchscreen Operation
Chapter 7
67
External versus Internal Actinic Light
The main panel of the window Actinic Light List (Fig. 36, page
67) indicates the light source selected for actinic illumination.
External Actinic Illumination is displayed when:
- An external lamp (2054-L External LED Source) is connected
to the SYNC port of the MINI-PAM and “PAM Mode (Act)” is
selected as operation mode (see 7.3.2.2, page 78).
Internal Actinic Illumination is displayed when:
- An external lamp (2054-L External LED Source) is connected
to the SYNC port of the MINI-PAM and “Ext. Panel Mode” is
selected as operation mode.
- An external lamp is not connected.
Actinic Light List
A
CAL
MENU
Side Panel
CAL
. Calibrate Actinic
Light List.
Go to previous win-
dow.
MENU
, go to Main
Menu.
1:
25
7:
285
2:
45
8:
420
3:
65
9:
625
4:
90
10:
820
5:
125
11:
1150
> 6:
190
12:
1500
External/Internal Actinic Light
2018-04-12 8:20:09
Main Panel
PAR target values (µmol m
-2
s
-1
) of the 12 settings for actinic light intensity.
The PAR target values are valid for an actinic light factor of 1.00 (see menu “Light
Sources” Fig. 41, page 77). The “>” indicates the currently selected PAR level.