Waltron User Manual 101-024-B.1
3048 Dissolved Iron Analyzer
User Interface
Cycles ratio:
Programs the ratio of the analysis cycles versus the extra cycle (calibration cycle).
For example, 400 analysis to 1 extra cycle means that every 400 analysis cycles the analyzer will
perform 1 calibration cycle.
Cycle wait:
Sets the time between analysis cycles when the analyzer is in on-line mode.
Result alarm:
Programs the low / high value alarm. The alarm is expressed in % of full scale.
Output Signal:
Enables / disables the 4-20 mA output.
Datalog:
Enables / disables the internal data logger.
Sample Alarm:
Enables / disables the loss of sample alarm.
5.8.2
CALIBRATION MENU
Allows the user to display and set the value of the standard used for calibration and perform
manual blank and calibration cycles (refer to Section 6).
Figure 5.9: User interface at the CALIBRATION Menu.