TECHNICAL SPECIFICATIONS
42
TC6XX-TM6XX Series
G.2 Thermocouple emission function (TC621)
Rated maximum voltage in common mode: 60 VDC or VAC.
G.2.1
Constant voltage
The voltage emission is made by configuring the device as follows:
Thermocouple: indifferent.
Unit: mV.
CSF: OFF.
Gauge
Scope of measurement
Resolution (min)
Precision / 1 year
80 mV
-9.5 mV / 80 mV
1 µV
0.020% R + 3 µV
Temperature coefficient < 10 ppm R /°C from 0°C to 18°C and from 28°C to 50 °C.
Internal resistance: ≤ 1 Ω.
G.2.2
Temperature by Thermocouple
Type of sensors:
- Standardized in accordance with CEI 584-1/1995 (Thermocouples K. T. J. E. S. B. N).
- In accordance with Din 43710 (Thermocouples U and L).
- In accordance with the HOSKINS table (Thermocouple C).
- In accordance with the ENGELHARD table (Platinum Thermocouple).
- Standardized in accordance with ASTM E 988-96 /2002 (Thermocouple D W3Re/W25Re)
- Standardized in accordance with ASTM E 1751-00 /2000 (Thermocouple G W/W26Re)