7  Calibration 

As with all electronic RTD thermometers, there are two main components to the 
system.  The first component is the electronics, which measures the resistance of 
the sensing element and then converts this resistance value to a temperature 
indication.  The second component is the probe, also referred to as the sensor.  
The probes sensor resistance changes as its temperature changes.  This change 
is in conformance to the DIN EN 60751 Class A standard.  As in all manufactured 
goods, there are slight variations in the finished parts.  A key variable of RTD 
sensors is its resistance at 0

°

C, referred to as “RØ” value.  Once this value is 

measured, it may be programmed into the meter to correct the temperature 
conversion algorithm.  Programming and/or calibration require the use of the 
DSTCAL software package.  This package includes the USB cable, USB/DST 
Interface Box and DST calibration cable.  DSTCAL programming software 
includes the ability to set the Sample Rate, RØ value and allows calibration of the 
meter and/or calibration of the system (calibrates meter with probe’s R0 value).  
The DST500-FM uses the following methods for calibration.  For details of the 
calibration, see the DSTCAL software manual, Wahl P/N W1037. 
 
7.1  Ohm Meter calibration – Ohm Meter calibration is a 2-point calibration, 

which calibrates the DST500-FM electronics.  It requires a NIST traceable 
precision resistance source with a known accuracy of 

±

 0.03 ohms at 50.00 

and 280.00 ohms.   Resistors used should have a temperature coefficient of 
<5.0-ppm. 

 
7.2  Probe (system) calibration - Probe calibration is a single point calibration 

which is performed by placing the probe into an ice bath and following the 
on screen prompts of the DSTCAL software.  The DST500-FM and 
software will measure the probes RØ value and program it into the 
DST500-FM’s memory.  The system is then calibrated.  After system 
calibration is completed, the calibration should be checked at the 
temperatures of interest to the end user, using calibrated accurate 
standards and calibration baths. 

 
7.2.1  DST500-FM calibration adjustment - Calibration is accomplished via the 

DSTCAL software package and associated hardware interface.  Method 1 
may also be performed with the DSTPROG software.  Optional cable 
extensions are available for convenience in calibrating fixed probes.   

 
7.2.2  The two methods of calibrating a meter with probe as a system are: 
 
7.2.3  Method 1 

The first method uses an ice bath and high accuracy ohmmeter 
along with the DSTCAL software. 
For fixed probes: 
1) Disconnect the probe from the meter and place the probe in an 
ice bath or 0

°

C circulating bath.  Allow stabilization. 

Содержание DST500-FM

Страница 1: ...FM approval Wahl Instruments Inc 234 Old Weaverville Road Asheville NC 28804 Toll Free 800 421 2853 Phone 828 658 3131 Fax 828 658 0728 www palmerwahl com Intrinsically Safe Sécurité Intrinsèque IS CL I II III DIV 1 GP ABCDEFG T4A Ta 70 C Type 4X CL I II III DIV 2 GP ABCDEFG T4A Ta 70 C Type 4X ...

Страница 2: ...____________________________________ 9 8 Installation Control Drawing _________________________________________ 11 9 Specifications _____________________________________________________ 12 10 Service___________________________________________________________ 12 WARRANTY CALIBRATION REGISTRATION at www palmerwahl com register Registration is fast and easy In about a minute you can have your product...

Страница 3: ...film platinum RTD sensor per DIN EN 60751 Class A Hazardous Location approval by FM Approvals as defined in Section 3 of this manual High accuracy 24 bit Delta Sigma Analog Digital Converter Probe Error checking Checks for open wire open sensor shorted sensor incorrect wiring Programmable RØ Allows programming of the sensing probes specific RØ value into the meter for accurate temperature calibrat...

Страница 4: ...nclosure or the probe is grounded to building or earth ground Warning Substitution of the following components may impair suitability for Division 2 Battery Wahl P N DSA3060 Manufacturer Tadiran model TL 2200 3 3 Avertissements Français Avertissement S il vous plaît respecter les avertissements suivants Avertissement La batterie doit être changée en dehors d une zone classifiée Avertissement Risqu...

Страница 5: ...ional check that the battery is installed properly by following the battery installation procedure section 5 of this manual 4 1 Meter Mounting Fixed probes may be mounted by the threaded fitting or sanitary clamp into the process Apply thread sealing compound or Teflon tape to the threaded fitting as required Units with a swivel nut fitting may be adjusted during installation for best viewing angl...

Страница 6: ... small ferrite around the 4 probe wires and the green yellow ground wire as close to the ground wire spade terminal as possible Make certain not to pinch the probe wires in the ferrite Route the 4 probe wires through the solid ferrite and loop back around through the ferrite 2 more times for a total of 3 passes through the ferrite Connect wires to PCB plug P1 per probe wiring table 4 4 Grounding 4...

Страница 7: ... approximately 1 2 weeks of battery life left The battery should be replaced at this time As the battery continues to drain will appear indicating the battery is too low for an accurate measurement The battery must be replaced at this time Battery life will vary dependent on Display Update Rate Setting as shown in the Battery Life Table below 5 1 For installation or replacement loosen the four Phi...

Страница 8: ...s set to 8 for a 2 second update 6 3 Error Codes During normal operation the DST500 FM continually performs diagnostic testing on the sensor lines Errors are indicated by the following error codes Error Code Description HI Reading is above meters usable range LO Reading is below meters usable range or sensor is shorted CbL1 Indicates cable 1 is open J1 pin 1 CbL2 Indicates cable 2 is open J1 pin 2...

Страница 9: ...ware manual Wahl P N W1037 7 1 Ohm Meter calibration Ohm Meter calibration is a 2 point calibration which calibrates the DST500 FM electronics It requires a NIST traceable precision resistance source with a known accuracy of 0 03 ohms at 50 00 and 280 00 ohms Resistors used should have a temperature coefficient of 5 0 ppm 7 2 Probe system calibration Probe calibration is a single point calibration...

Страница 10: ...the probe connector and the meter PCB connector This allows the meter to be placed on a bench rather than dangling above the ice bath Place the probe in an ice bath Allow stabilization 2 Connect the meter to the PC via the 13506 Interface Module 3 Select and run the Probe system Calibration program and follow the on screen prompts A running average of R0 is displayed in the R0 box Press Capture R0...

Страница 11: ...DST500 FM USER MANUAL 11 of 12 8 Installation Control Drawing ...

Страница 12: ...echnical support contact Wahl Instruments Inc 234 Old Weaverville Road Asheville NC 28804 Ph 800 421 2853 US only 828 658 3131 Fax 828 658 0728 Web www palmerwahl com 234 Old Weaverville Road Asheville NC 28804 800 421 2853 828 658 3131 828 658 0728 www palmerwahl com info palmerwahl com ...

Отзывы: