3.0  PROBES
Probes compatible with the 392 HEAT-Prober are constructed utilizing a wire
wound Platinum RTD sensor with an R0 of 100 Ohms wired in a 3-wire con-
figuration to compensate for lead length effects. To provide probe inter-
changeability at 0 degrees C, a balancing resistor is added to the sensor cir-
cuit and trimmed for a total resistance of 100.94 Ohms at 0°C. (See Figure
3-2)

The High Performance HEAT-Prober, Model 392HP, utilizes probes similar to
the 392 but incorporating an additional calibration resistor that provides
probe interchangeability at all temperatures.

3.1 Interchangeability
Probes are individually balanced to 100.94 Ohms at 0°C to compensate for
variations in different probe configurations. This guarantees probe inter-
changeability at 0°C. In addition, each probe is tested at a minimum of one
higher temperature. This calibration data is recorded on a calibration tag
attached to the probe. (See Figure 3-1)

Probe Interchangeability (Immersion)

Ice Point

±0.3°C (±0.5°F)—(392HP ±0.2°F)

100°C

±0.6°C (+1.0°F)

200°C

±1.1°C (±2.0°F)

Surface probes are tested on surface calibration plates especially built to
replicate the end-use. The variance is somewhat larger as follows:

100°C

±1.1°C (±2.0°F)

200°C

±2.2°C (±4.0°F)

Figure 3-2.  Probe Circuit NOTE: Platinum sensor element manufacturers
quote 0.1% tolerance at 0°C, or 0.1 ohms on a 100 ohm sensor. This results
in 0.3°C variance. The tolerance at higher temperatures is greater than 0.1%
in accordance with the following formula as specified by DIN 43760:
Tolerance = ±.3°C + .0054t (°C)
Thus at 200°C, the expected sensor error could be 0.3°C + .0054 (200) =
±1.38°C.

3.2  Temperature Sensors in Use
An understanding of heat-transfer is needed for accurate field use of temper-
ature sensors. The sensor measures its own temperature, therefore, to meas-
ure a body temperature it requires good contact with that body, or in the case
of liquids or gases good immersion and mixing.

A portable probe temperature sensor must have the following characteristics:

(a) a low thermal mass (product of its specific heat and weight);
(b) a large area for surface contact to permit heat-transfer between the

object being measured and the probe sensor,

(c) a minimum of heat conduction away from the sensor which will

reduce its temperature below that of the object being measured and create an
inaccuracy.

Surface probes are designed with the above three conditions as criteria, and
also include an insulating handle for your safety and convenience. Each
probe is assembled with a plastic handle made of high quality Delrin

®

which:

(a) Thermally isolates the hand from the probe, preventing hand heat

from reaching the sensor, and sensor heat, in the case of high temperature
measurement, from burning the hand.

(b) Electrically isolating the hand from the metal probes and therefore,

potentially, from a voltage on the surface to be measured.

The handle is attached to a tangle-free retractable cord which extends 6' and
retracts to 12" (See Figure 2-1). Each Probe is labeled with its part number
and type.
For surface measurements, to obtain best time responses and accuracy, two
things are necessary;

(a)  Good surface contact with all of the probe surface. Make sure of

this by holding it squarely on the surface. The 121 and 145 surface probes
are specially designed with a spring loaded tip to provide compliance with the
surface.

(b)  A rough surface gives poor heat conductance to the measuring

probe. This can be overcome by applying a thin layer of heat conducting sili-
cone paste between the measuring point and the probe tip. The paste short-
ens response time up to 50%.

In liquids, move the immersion probe around the mix to get good contact. In
air, put the probe in the vicinity of a stream to get good velocity, otherwise
move the probe back and forth to induce velocity across the tip.

Содержание 392M

Страница 1: ...Prober PLATINUM RTD THERMOMETERS Models 392HP 392E 392E S 392EX 392EX S 392EVX 392M 392M S 392MX 392MX S 392VX 360X and 700M INSTRUCTION MANUAL Copyright 2004 Wahl Instruments Inc Printed in USA W 392 Rev M ...

Страница 2: ...s authorized to assume for Wahl any liablility except as set forth above Warranty Temperature Probes and Meter Accessories Wahl warrants its Hand Held Meter Temperature Probes to be free of defects in workmanship or material Due to the environmental nature of their use probes cannot be replaced without cost unless returned unused within a 30 day period Wahl assumes no liability for damages of any ...

Страница 3: ...y in all light conditions To minimize power drain the display is multiplexed The LCD display is 0 4 high with excellent contrast A large portion of the circuit is designed with low power C MOS components The result is a design which provides a low power drain operation which allows operation for 8 hours per charge for the 392E 392EX 392M and 392MX meters Battery life is 50 hours for the replaceabl...

Страница 4: ...e to conserve battery life 2 1 Meter Operation The HEAT Prober instrument has an accurate measuring range as stated on the front label The probes however may not be capable of this entire range of use Wahl probes are typically capable of 40 C to 500 C or 40 F to 900 F The digital readout will show resolution to 0 1 increments from 180 0 C to 190 C or 290 F to 375 F At approximately 190 C or 375 F ...

Страница 5: ...able of 50 hours continuous operation 2 4 1 Recharging To recharge the batteries use the recharger supplied Plug in directly to 115VAC 60 cycle wall plug and insert jack into socket on left side of HEAT Prober For overseas 220 volt lines use a Franzus converter adapter with the recharger supplied or request a 220 volt recharger 2 4 2 Continuous Operation The 392HP 392E 392EX 392M and 392MX HEAT Pr...

Страница 6: ...a body temperature it requires good contact with that body or in the case of liquids or gases good immersion and mixing A portable probe temperature sensor must have the following characteristics a a low thermal mass product of its specific heat and weight b a large area for surface contact to permit heat transfer between the object being measured and the probe sensor c a minimum of heat conductio...

Страница 7: ...thane Propane E Metal Dust F Coal Dust and G Grain Dust All 392 Models Available in F C or Dual Scale add F C or D to Model Number Dual Scale not available on S Models 392 SERIES METER SPECIFICATIONS Model 392M 392MX 392M S 392MX S 392VX Range 60 to 752 F 51 to 400 C 60 to 752 F 51 to 400 C Digital Display 392M 392M S Bright 0 33 high LED Red 0 4 high LCD 392MX 392MX S Red 0 4 high LCD Resolution ...

Страница 8: ...igure 5 1 392M Platinum Heat Prober with Wahl Calibrator Model Number CAL392HT 6 0 SET UP AND CALIBRATION 6 1 Equipment Needed A Digital voltmeter Fluke 41 2 digit Model 8600A or equivalent B Wahl Platinum Calibrator Model Number CAL392HT or CAL392LT or General Radio decade box Model 1433 1 or equivalent with standard coil cord Model 10434 01 or 10434 02 cable as used with Wahl probes C Power supp...

Страница 9: ...t a quick and accurate calibration can be achieved 6 6 1 Setup and Calculations The decade resistor box may be used as the source to perform a standard calibration A coiled cord cable with connector for the Wahl probe should be used Model Number 10434 01 This has an approximate resistance of 0 36 ohms The actual resistance should be determined for your setup and this value used in the calculations...

Страница 10: ...and 392EVX Models 392HP 392M and 392E Use the Circuit Board Diagrams on these pages to locate potentiometer placement of Zero Adjustment and Span Adjustment Note that placement varies with Model See page 28 for Model 700M Board Layout ...

Страница 11: ...e entire temperature range This also results in superior system accuracies at all temperatures The HP option utilizes a Noise Rejection program which minimizes transient noise effects In this version a special 4 pin receptacle is installed at the probe connection and the four pins are wired to the printed circuit board from the probe con nector side as shown in Figure 8 1 A resistor with a nominal...

Страница 12: ...the program Details of the components will be discussed in this section 9 2 Program Flow When Power is turned on the program will start executing instructions from address 000 the location of the top level program labeled Main The first routine called is Initialize This is a house keeping routine which establishes the starting conditions The data memory RAM is cleared the input output ports are co...

Страница 13: ...rage be crammed with new data Additionally if over range is displayed indicating on open probe lead or removed probe conditions are set for the next entry into Gaincheck to cause the compensation reading to be crammed and also that rapid updates to it are to be made on following passes This is done to accommodate changes in probes by the user so that the display will repre sent the actual probe re...

Страница 14: ... 58MHz square wave with levels from 0 to 5 volts Replace U6 If still no display go to Step 9 Replace U4 If still no dis play go to Step 7 4 Check voltage at Q4 collector It should be equal to the battery voltage or applied voltage from test power supply Proceed to next step Check for open trace from Pin 5 to Q4 C 5 Check voltage on junction of R30 and CR5 cathode Should be 5 000 005V Proceed to ne...

Страница 15: ...16 Return unit to factory for repair NOTE Does not apply to 392MX and 392M with Serial Numbers before N120 11 0 ACCESSORIES 392 SERIES CALIBRATORS Model Number Usable With Meter Calibration Points CAL392 HT 392E 392EX 392EVX 392M 392MX 392VX and 700M 58 F 32 F 212 F 284 F 752 F 50 C 0 C 100 C 140 400 C CAL392HP HT 392HP only CAL392 LT 392E 392EX 392EVX 392M 392MX 392VX Ice Point to Boil Calibrator...

Страница 16: ...s the same as the 392M which is a recharge able nickel cadmium battery pack composed of five AA size cells with a com bined voltage of 6 0 volts and capacity rating of 65 amp hours The unit is supplied with a recharger which can be plugged in at any time to maintain the charge of the battery pack Low battery condition is displayed as L in the left hand side of the display The 700M is available onl...

Страница 17: ...50 plug connector 3 Set resistance to 29 490 ohms and adjust R5 for a display read ing of 0 00 C 32 F 4 Set the decade box to a resistance of 2760 ohms and adjust R13 for a display reading of 60 00 C 140 00 F 5 Set the decade box to a resistance of 12 260 ohms and check the unit for a correct display reading of 20 00 C or 68 00 F 6 Repeat steps 3 4 and 5 as necessary for best calibration 14 0 392M...

Страница 18: ...d to next step 8 Measure voltage at pin 6 of wire harness It should be equal to the battery or power supply voltage Proceed to next step Check wire harness for broken wires 9 Replace Q1 and repeat Step 7 Done Go to Step 19 10 Unit con stantly read OL Measure voltage at U4 28 with DVM It should be 4 3 0 2V Go to step 12 Proceed to next step 11 Check voltage at U3 8 It should be 5 0V 0 25V Replace U...

Страница 19: ...u follow up reminders about recalibration dates 16 0 RETURN FOR REPAIR CALIBRATION SERVICE Warranty Repair Information Please contact Wahl Customer Service at 800 421 2853 for a Return Merchandise Authorization number RMA before returning your instrument to the factory Shipping Material being returned should be packed well preferably in the original foam shipping container Ship to WAHL INSTRUMENTS...

Отзывы: