DC current measurement

 

CAUTION:  Do not make 20 A current measurements for longer  

 

than 30 seconds. Exceeding 30 seconds may cause damage to the 

 

 meter and/or the test leads.

1. 

Insert black test lead banana plug into the negative COM jack.

2. 

For current measurements up to 6000µA DC, set the function switch to 

the µA position and insert the red test lead banana plug into the µA/

mA jack

3. 

For current measurements up to 600mA DC, set the function switch to 

the mA position and insert the red test lead banana plug into the µA/

mA jack

4. 

For current measurements up to 20A DC, set the function switch to the 

10A/HZ/% position and insert the red test lead banana plug into the 

10A jack

5. 

Press the MODE button to indicate “DC” on the display

6. 

Remove power from the circuit under test,  

then open up the circuit at the point  

where you wish to measure current

7. 

Touch the black test probe tip to the  

negative side of the circuit

8. 

Touch the red test probe tip to the  

positive side of the circuit

9. 

Apply power to the circuit

10.  Read the current in the display

Содержание Nieaf-Smitt NI 74

Страница 1: ...Nieaf Smitt is a brand name of Wabtec NETHERLANDS NI 74 True RMS multimeter Handleiding Manual...

Страница 2: ...ic in any form or manner either electronically mechanically by photocopying recording or in any manner without prior written consent from Wabtec Netherlands B V This also applies to accompanying drawi...

Страница 3: ...14 4 5 Weerstand meting 15 4 6 Continuity check 16 4 7 Diode test 17 4 8 Temperatuur meting 18 4 9 Capaciteit meting 19 4 9 Frequentie duty cycle meting 20 5 Functies 5 1 Auto handmatig bereik 20 5 2...

Страница 4: ...n kunnen leiden Symbolen op de multimeter Dit symbool naast een ander symbool ingang of regelorgaan geeft dat de gebruiker de handleiding moet raadplegen om persoonlijk letsel of schade aan de meter t...

Страница 5: ...een meting Verricht nooit een spanningsmeting met een meetsnoer aange sloten op de A invoer Controleer de werking van de multimeter door het meten van een bekende spanning Laat bij twijfel de multimet...

Страница 6: ...t een hoge spanning voor dat men weerstand continu teit doorgang diodes of capaciteiten in dit circuit gaat meten Gebruik de multimeter niet in de buurt van explosieve gassen of dampen Reduceer het ge...

Страница 7: ...ANNINGSCATEGORIE I is apparatuur voor aansluiting op een circuit welke maatregelen worden getroffen om de transient overspan ningen te beperken tot een veilig laag niveau Overspanning categorie II App...

Страница 8: ...erstand Duty cycle Temperatuur Kenmerken True RMS Stroommeting tot 20 A AC DC Capaciteit Doorgangs diodetest IP 67 waterdicht Min max hold Automatische uitschakeling Verlicht display Schermresolutie 6...

Страница 9: ...orpaneel 1 LCD display 2 Hz toets 3 RANGE toets 4 MODE toets 5 Functie schakelaar 6 mA A en 10A ingangsbus 7 COM ingangsbus 8 Positieve ingangsbus 9 Achtergrondverlichting en Hold toets 10 MAX MIN toe...

Страница 10: ...amps cap m milli 10 3 volt amp A Amps k kilo 103 ohms F Farads capaciteit M mega 106 ohms Ohms Hz Hertz frequentie V Volt REL Relative AC Alternatieve stroom DC Gelijkspanning HOLD Display hold F Gra...

Страница 11: ...4 1 DC voltage meting WAARSCHUWING Meet geen DC voltages als er een motor in het circuit aan of wordt uitgeschakeld De grote spanning schommelingen kunnen mogelijk de meter beschadigen 1 Plaats de fun...

Страница 12: ...geen AC spanning als er een motor in het circuit aan of wordt uitgeschakeld Grote spanning schommelingen kunnen mogelijk de meter beschadigen Plaats de functieschakelaar op de groene VAC Hz positie 7...

Страница 13: ...ke laar op de mA positie en plaats het rode meetsnoer stekker in de A mA stekker 4 Voor stroom metingen boven de 20 A DC plaats de functie schakelaar op de 10 A HZ positie en plaatst het rode meetsnoe...

Страница 14: ...positieve A mA stekker 4 Voor een stroom meting van 20 A AC schakel de functie schakelaar in op de 10 A Hz positie en plaats het rode meetsnoer met banaanstekker in de 10 A stekker 5 Druk de MODE kno...

Страница 15: ...aar op de groene positie 2 Plaats het zwarte meetsnoer met banaanstekker in de negatieve COM stekker 3 Plaats het rode meetsnoer met banaanstekker in de positieve stekker 4 Druk de MODUS knop voor de...

Страница 16: ...meetsnoer met banaanstekker in de negatieve COM stekker 3 Plaats het rode meetsnoer met banaanstekker in de positieve stekker 4 Gebruik de MODUS knop voor de indicatie van X en op het scherm weergegev...

Страница 17: ...M stekker 3 Plaats het rode meetsnoer met banaanstekker in de positieve V stekker 4 Gebruik de MODE knop om de X indicator in het scherm weer te geven 5 Raak met de uiteinden van de testsonde de diode...

Страница 18: ...ratuur probe tegen het onderdeel wat u wil meten 5 De probe moet 30 seconden in contact blijven om de test uit te voeren en de temperatuur te stabiliseren 6 Lees de temperatuur af in het scherm Opmerk...

Страница 19: ...AP X positie 2 Plaats het zwarte meetsnoer met banaanstekker in de negatieve COM stekker 3 Plaats het rode meetsnoer met banaanstekker in de V positie 4 Druk de MODUS knop in om de indicator nF te lat...

Страница 20: ...le of metingen op het scherm 5 Functions 5 1 Auto handmatig bereik Wanneer de meter voor het eerst wordt aangezet wordt automatische bereik selectie ingesteld Dit automatische selectiesysteem kiest he...

Страница 21: ...ter zal weergeven en Hold de minimale waarde en zal alleen updaten wanneer er nieuwe min is gemeten 6 Op de MAX MIN modus uit te gaan druk en houd de MAX MIN toets in voor 2 seconden 5 3 RELATIVE modu...

Страница 22: ...hakelen 5 6 Auto Power Off De meter zal automatische uitgeschakelen wanneer hij 15 min buitengebruik is Om de meter na een automatische uitschakeling in te schakelen draai de MODE knop naar ON 5 7 Leg...

Страница 23: ...met een vochtige doek gebruik geen chemische producten schoonmakende oplosmiddelen of afwasmiddel 5 GEBRUIK ALLEEN VERSE BATTERIJEN VAN DE GEADVISEERDE GROOTTE EN HET TYPE Verwijder oude of zwakke bat...

Страница 24: ...ed ge nstalleerd zijn 6 2 Zekering vervangen WAARSCHUWING Om elektrische schok te vermijden ontkoppel de meetsnoeren van mogelijke stroom bron voordat U de zekering deksel verwijderd 1 Ontkoppel de me...

Страница 25: ...1000 V snel slag voor de 600 mA bereik SIBA 70 172 40 10 A 1000 V snel slag voor 20A bereik SIBA 50 199 06 7 Plaats en bevestig de achterkant en de batterijdeksel WAARSCHUWING Om een elektrische schok...

Страница 26: ...V 1 mV 0 8 4 d 60 00 V 10 mV 600 0 V 100 mV 1000 V 1 V 1 2 4 d Alle AC voltage ranges zijn specifiek van 5 of ranges tot 100 Invoer impedantie 10 M AC Response 50 Hz 400 Hz Maximum invoer 1000 VDC of...

Страница 27: ...0 Hz 400 Hz Maximum invoer 6000 uASC rms bij uA bereik 800 mAAC rms bij mA bereik 10 AAC rms bij 10 A bereik Functie Bereik Resolutie Nauwkeurigheid Weerstand auto ranging 600 0 0 1 0 5 4 d 6 00 K 1 6...

Страница 28: ...urigheid Duty cycle 0 1 99 9 0 1 1 2 2 d Pulse width 100 us 100 ms Frequentie width 5 Hz 150 kHz Gevoeligheid 0 5 V RMS Overbelasting bescherming 1000 VDC of AC rms Functie Bereik Resolutie Nauwkeurig...

Страница 29: ...rue RMS ACV bandbreedte 50 Hz 1000 Hz Crest factor 3 bij volledige schaal tot 500 V liniare verlaging tot 1 5 at 1000 V Display 60 00 count achtergrond verlichting met bargraph Overrange indication OL...

Страница 30: ...x 50 mm Veiligheid Deze meter is bedoeld en beschermd voor binnenge bruik tegen de gebruikers door dubbele isolatie per EN61010 1 CAT III 1000 V CAT IV 600 V Verontreiniging Graad 2 Garantie Wabtec N...

Страница 31: ...urement 42 4 5 Resistance measurement 43 4 6 Continuity check 44 4 7 Diode test 45 4 8 Temperature measurement 46 4 9 Capacitance measurement 47 4 9 Frequency duty cycle measurement 48 5 Functions 5 1...

Страница 32: ...tructions to avoid personal injury or damage to the meter This WARNING symbol indicates a potentially hazardous situation which if not avoided could result in death or serious injury This CAUTION symb...

Страница 33: ...between terminals or between any terminal and earth ground Do no apply more than the rated voltage as marked on the multimeter between terminals or between any terminal and earth ground Only replace...

Страница 34: ...input jack exceeds 1000 V above earth ground Never connect the meter leads across a voltage source while the func tion switch is in the current resistance or diode mode Doing so can damage the meter F...

Страница 35: ...s equipment for connection to circuits in which measures are taken to limit the transient overvoltages to an appropriate low level Examples include protected electronic circuits Overvoltage category I...

Страница 36: ...rent AC DC Frequency Resistance Duty cycle Temperature C F Features True RMS Current measurement 20 A AC DC Capacity Continuity diodetest IP 67 waterproof Min max hold Automatic power off Backlight Re...

Страница 37: ...ensor 6 Manual Front panel 1 6000 count LCD display 2 Hz button 3 RANGE button 4 MODE button 5 Function switch 6 mA A and 10A input jacks 7 COM input jack 8 Positive input jack 9 Backlight and Hold bu...

Страница 38: ...10 6 amps cap m milli 10 3 volt amp A Amps k kilo 103 ohms F Farads capacitance M mega 106 ohms Ohms Hz Hertz frequency V Volts REL Relative AC Alternating current DC Direct current HOLD Display hold...

Страница 39: ...Change to a higher range 4 1 DC voltage measurements CAUTION Do not measure DC voltages if a motor on the circuit is being switched ON or OFF Large voltage surges may occur that can damage the meter...

Страница 40: ...present CAUTION Do not measure AC voltages if a motor on the circuit is being switched ON or OFF Large voltage surges may occur that can damage the meter 1 Set the function switch to the green VAC Hz...

Страница 41: ...up to 600mA DC set the function switch to the mA position and insert the red test lead banana plug into the A mA jack 4 For current measurements up to 20A DC set the function switch to the 10A HZ posi...

Страница 42: ...plug into the A mA jack 4 For current measurements up to 20A AC set the function switch to the 10A HZ position and insert the red test lead banana plug into the 10 A jack 5 Press the MODE button to i...

Страница 43: ...nction switch to the green position 2 Insert the black test lead banana plug into the negative COM jack 3 Insert the red test lead banana plug into the positive jack 4 Press the MODE button to indicat...

Страница 44: ...sition 2 Insert the black lead banana plug into the negative COM jack 3 Insert the red test lead banana plug into the positive jack 4 Press the MODE button to indicate and on the display 5 Touch the t...

Страница 45: ...and the red test lead banana plug into the positive V jack 3 Press the MODE button to indicate and V on the display 4 Touch the test probes to the diode under test 5 Forward voltage will typically in...

Страница 46: ...temperature probe head to the part whose temperature you wish to measure 5 Keep the probe touching the part under test until the reading stabilizes about 30 seconds 6 Read the temperature in the displ...

Страница 47: ...y function switch to the green CAP position 2 Insert the black test lead banana plug into the negative COM jack 3 Insert the red test lead banana plug into the positive V jack 4 Press the MODE button...

Страница 48: ...display 5 Functions 5 1 Auto manual range When the meter is first turned on it automatically goes into Autoranging This automatically selects the best range for the measurements being made and is gene...

Страница 49: ...ding and will update only when a new min occurs 6 To exit MAX MIN mode press and hold the MAX MIN key for 2 seconds 5 3 RELATIVE mode The relative measurement feature allows you to make measurements r...

Страница 50: ...feature hold down the MODE button and turn the meter on 5 7 Low battery indication The icon will appear in the lower left conner of the display when the battery voltage becomes low Replace the battery...

Страница 51: ...ak batteries so they do not leak and damage the unit 6 If the meter is to be stored for a long period of time the batteries should be removed to prevent damage to the unit 6 1 Battery installation WAR...

Страница 52: ...emove the protective rubber holster 3 Remove the battery cover two B screws and the battery 4 Remove the six A screws securing the rear cover 5 Gently remove the old fuse and install the new fuse into...

Страница 53: ...0 V 1 mV 0 8 4 d 60 00 V 10 mV 600 0 V 100 mV 1000 V 1 V 1 2 4 d All AC voltage ranges are specified from 5 of range to 100 of range Input Impedance 10 M AC Response 50 Hz 400 Hz Maximum Input 1000 VD...

Страница 54: ...esponse 50 Hz 400 Hz Maximum input 6000 uASC rms on uA 800 mAAC rms on mA 10 AAC rms on 10 A range Function Range Resolution Accuracy Resistance auto ranging 600 0 0 1 0 5 4 d 6 00 K 1 60 00 K 10 0 5...

Страница 55: ...olution Accuracy Duty cycle 0 1 99 9 0 1 1 2 2 d Pulse width 100 us 100 ms Frequency width 5 Hz 150 kHz Sensitivity 0 5 V RMS Overload protection 1000 VDC or AC rms Function Range Resolution Accuracy...

Страница 56: ...True RMS ACV bandwidth 50 Hz 1000 Hz Crest factor 3 at full scale up to 500 V decreasing linearly to 1 5 at 1000 V Display 60 00 count backlit liquid crystal with bargraph Overrange indication OL is d...

Страница 57: ...se and protected against the users by double insulation per EN61010 1 to Category IV 600 V and Category III 1000V Pollution Degree 2 The meter also meets UL 61010 1 2 Edition 2004 CAN CSA C22 2 No 610...

Страница 58: ...58 www nieaf smitt com...

Страница 59: ...59 www nieaf smitt com...

Страница 60: ...60 www nieaf smitt com...

Отзывы: