Replacing the battery 

WARNING: 

REPLACE BATTERY WITH ANY GENERAL PURPOSE OR HEAVY DUTY 9-VOLT CELL.  FOR 

LONGER LIFE USE DURACELL 1604, ENERGIZER NO. 522, RADIOSHACK NO. 23-653, 

EVEREADY NO. 1222, RAYOVAC NO. A1604. 

 
To replace the battery you will need to remove the bottom end of the ISOTHERM 2000's handle.  
It unscrews similar to a common flashlight.  Inside the handle, there is a 9V battery attached to a 
long connector with leads back into the handle.. 
 

1) Remove the battery from the handle 

 

2) Remove the battery connector from the battery 

 

3) Connect the new battery to the connector  

 

4) Put the battery back into the handle 

 

5) Replace the end of the handle 

 
Reel Version  
 

1) Remove the six screws from the face and remove  
2) Remove the battery from the battery holder  

 

3) Disconnect the battery  

 

4) Connect the new battery 

 

5) Put the battery back in the battery holder  

 

6) Replace the face and screws (be sure that o-ring is seated properly) 

 
 
Replacing the cable 

WARNING: 

SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY 

AVERTISSEMENT: 

LA SUBSTITUTION DE COMPOSANTS PEUT COMPROMETTRE LA SÉCURITÉ 

INTRINSÉQUE 

 
The probe and cable assembly on the ISOTHERM 2000 is designed to be replaced with minimal 
effort.  Follow these easy steps: 

1) Obtain a good replacement cable (new or rebuilt from W.L. Walker Co., Inc.) 
2) Unwrap the old cable from the ISOTHERM 
3) Unscrew the two screws that attach the end of the cable to the ISOTHERM 
4) Remove the old cable 
5) Attach the new cable to the screws vacated by the old cable 
6) Replace the screws removed in step 3. 
7) Test the connection by turning the unit on.  If the unit displays the current temperature 

correctly then proceed to step 8.  Otherwise, check the replacement cable to determine 
that is attached to the ISOTHERM 2000 properly. 

8)  Attach the ground clip to the probe and turn on the ISOTHERM again.  If the ISOTHERM 

still reads the correct temperature, then you are done installing the new cable.  If the 
ISOTHERM now reads SHRT (for short), then you have attached the two ends of the 
cable backwards.  Swap there positions and restart with step 5. 

9) Wind the new cable onto the ISOTHERM 

 
 
 
 
 
 
 

Содержание ISOTHERM 2000

Страница 1: ...Using the ISOTHERM 2000 W L Walker Co Inc Tulsa Oklahoma 888 397 9255...

Страница 2: ...s DEPRESSED The display will show 8 8 8 8 8 8 8 This demonstrates that all of the display segments are still working After the initial display the ISOTHERM 2000 will switch to displaying the current t...

Страница 3: ...SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY AVERTISSEMENT LA SUBSTITUTION DE COMPOSANTS PEUT COMPROMETTRE LA S CURIT INTRINS QUE The probe and cable assembly on the ISOTHERM 2000 is designe...

Страница 4: ...se the patch cable that is provided with the calibration tool Using the Calibration Tool 2000 While in the calibration mode you can adjust the temperature up and down with the Calibration Tool 2000 Pr...

Страница 5: ...ture If you don t wait 5 minutes or more you won t get a good calibration Adjust the ISOTHERM s temperature until it matches the bath temperature See Using The Calibration Tool 2000 above to find out...

Страница 6: ...nt to replace your battery immediately to prevent down time for the unit The OPEn display could appear in the following cases 1 There is no Probe Cable assembly attached to the ISOTHERM 2 The Probe Ca...

Страница 7: ...o the ISOTHERM the ISOTHERM plugged into the ISOTHERM 2000 2000 Using incorrect procedure Review the ISOTHERM MARK calibration instructions I can t get a good Temperature hasn t stabilized Wait about...

Страница 8: ...________________________________________________________________ ________________________________________________________________ WALKER LAB USE Date Tulsa ____ ____ ____ by __________ Description of...

Отзывы: