background image

Figure 1 - Back View of ARL Module

I. SETTING UP the PTR-200

A. The following steps will show you how to set up your PTR-200. (Refer to Fig.1)

1.

Attach the Integrated Module Controller (VYT-200-C), by removing the four mounting screws and washers from the  

back of the PTR-200.  Place the PTR-200 and Controller on a flat work surface and secure the controller bracket to 
the back of the PTR-200.  Make sure the feet on the bottom of the bracket sit flush on the work surface.  

NOTE:

You can set the tension of the tilt display by adjusting the left- hand pivot screw.  A 5/64" allen wrench is provided

in the tool kit.

2

.  Connect the interface cable from the bottom of the Controller to the back of the PTR-200. 

3.

An RS-232 serial input is provided for direct PC control.  

Note:

The Integrated Module Controller and the PC 

serial interface should not be connected simultaneously - possible damage to the controller could occur. 

4.

Connect the DC power cable from the external power supply to the back of the PTR-200.  Plug in the power supply 

(100-120 / 200-240 VAC).

5.

Plug in the external vacuum pump.  To connect the vacuum line, insert the silver quick connect fitting into the 

lower fitting on the pump.  Connect the other end of the tubing to the brass barbed fitting on the back of the 
PTR-200.  Connect the small "muffler" to the upper fitting of the vacuum pump.

6. 

The pneumatic gripper of the automated recoater requires a clean compressed air (or gas) supply of  80 to 120 psi.  

Use the fittings kit provided to connect the 1/8" OD quick connect fitting into the supply source.  Connect the 1/8"
OD pressure supply line between the source and the unit by pushing the tubing fully into the fittings.

PTR-200-ARL Quick Reference Guide   Version 1.2

Page 2

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Содержание PTR-200 Series

Страница 1: ...l service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE INSPECTION Remotely inspect equipment before purchasing with our interactive website at www instraview com LOOKING FOR MORE INFORMATION Visit us on the web at www artisantg com for more information on ...

Страница 2: ...R RE EF FE ER RE EN NC CE E G GU UI ID DE E PTR 200 ARL Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Страница 3: ...oftware may be copied photocopied reproduced translated or reduced to any electronic medium or machine readable form in whole or part without prior written consent of Vytran Corporation Copyright 1999 Vytran Corporation 1400 Campus Drive West Morganville NJ 07751 USA Phone 732 972 2880 800 421 8874 Fax 732 972 6229 PTR 200 ARL Quick Reference Guide Version 1 2 Artisan Technology Group Quality Inst...

Страница 4: ...ure that you have the following accessories 12 5V Power supply box with AC power cord DC power cable to connect the external power supply to the unit 1 8 O D pressure supply tubing and fittings kit for compressed air hook up VYT 200 C Integrated Module Controller with interface cable optional External vacuum pump with vacuum line optional Tool kit containing Recoat material DSM Desotech 950 200 Co...

Страница 5: ...erial interface should not be connected simultaneously possible damage to the controller could occur 4 Connect the DC power cable from the external power supply to the back of the PTR 200 Plug in the power supply 100 120 200 240 VAC 5 Plug in the external vacuum pump To connect the vacuum line insert the silver quick connect fitting into the lower fitting on the pump Connect the other end of the t...

Страница 6: ... B Power Up the PTR 200 1 Turn on the external power supply the external vacuum pump optional and the PTR 200 Fig 1 7 The system will go through an initialization process which takes from 15 90 seconds 2 After initialization you will be prompted to Press Start to Continue Touch the highlighted area of the screen to proceed to the Main Menu Note It may take 5 10 minutes for the controller to warm u...

Страница 7: ...2 Make sure to have cotton swabs and cleaning solution acetone or alcohol available prior to proceeding The recoat material must be collected as it comes out of the mold injection port Do not allow recoat material to run down the face of the mold and under the mold plate 3 From the TOOLS menu select the INJECT µl icon and using the numeric keypad enter in 50 for the injection volume You must injec...

Страница 8: ...ct rate is 1 8µl s Delay Time The amount of time in seconds between the end of the injection process and the start of the curing process lamps on Lamp Time The amount of time in seconds that the curing lamps are on Recommended values are preset at time of shipment Typical values for a 30mm long recoat strip length and an ARL Module with a 260mm mold are listed below Quantity l 1 2 Rate l s 0 2 Del...

Страница 9: ...as follows Diameters and length of the recoat in mm Select the Correct Mold Size for Your ARL Module 160 m 260 m 280 m 430 m 900 m 5 mm 0 07 0 20 0 25 0 86 4 00 10 mm 0 13 0 40 0 50 1 72 8 00 15 mm 0 20 0 60 0 75 2 58 12 00 20 mm 0 27 0 80 1 00 3 44 16 00 25 mm 0 34 1 00 1 25 4 30 20 00 30 mm 0 40 1 20 1 50 5 16 24 00 40 mm 0 54 1 60 2 00 6 88 32 00 Recoat length Vol l ________ ________ length of ...

Страница 10: ...Ramp Time The approximate time in seconds required to reach the peak tension Hold Time The amount of time in seconds that the peak load is held for Typical values for proof testing standard single mode fiber eg Corning SMF 28 are Units kpsi Fiber Dia 125 m Peak Tension 100 Ramp Time 5 seconds Hold Time 0 seconds 2 To change a parameter highlight the desired parameter type in the desired value and ...

Страница 11: ... or a soft wipe between the mold plates Use several sheets or fold several times to get multiple layers 3 Select the TOOLS icon and highlight the STORE command Select to close the mold and release the pneumatic pressure to the gripper 4 Turn off the vacuum pump if provided 5 Turn off the power switch on the back of the PTR 200 6 Turn off the power supply PTR 200 ARL Quick Reference Guide Version 1...

Страница 12: ...1400 Campus Drive West Morganville NJ 07751 USA Tel 732 972 2880 Fax 732 972 6229 s s s Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Страница 13: ...1400 Campus Drive West Morganville NJ 07751 USA Tel 732 972 2880 Fax 732 972 6229 s s s Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Страница 14: ...l service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE INSPECTION Remotely inspect equipment before purchasing with our interactive website at www instraview com LOOKING FOR MORE INFORMATION Visit us on the web at www artisantg com for more information on ...

Отзывы: