Measuring ranges, resolution, accuracy
Measuring
ranges,
resolution
Connection cable
Length
1.5 m
Diameter
4.3 mm
Smallest allowed
bend radius
Fixed installation: 20 mm
Flexible use: 60 mm
Plug type
Socket, 4 pins
Pressure resis-
tance
Sensor with connection
cable
IP 68 (2.5 x 10
5
Pa or 2.5 bar)
The IDP 761-C meets the requirements according to article 3(3) of the
directive, 97/23/EC ("pressure equipment directive").
Measurement con-
ditions
Conductivity measuring
range
1
S/cm ... 500 mS/cm
Temperature range
-5 ... 70 °C (100 °C) 23 ... 158 °F (212 °F)
Max. admissible overpres-
sure
IDP 761-C: 2.5 x 10
5
Pa (2.5 bar)
Minimum depth of immer-
sion
36 mm
Maximum depth of immer-
sion
(at temperature)
Whole cable up to 70 °C (158 °F)
Sensor shaft only (=120 mm) up to 100 °C
(212 °F)
Operating position
Any
Storing conditions
Recommended storing
method
In air
Storage temperature
0 ... 50 °C (32 ... 122 °F)
Characteristics
when delivered
Temperature responding
behavior
t
99
(99 % of the final value display after)
< 20 s
Accuracy of the temperature
sensor
± 0.2 K
Measured parameter
Measuring range
Resolution
ӕ
[µS/cm]
0.0 ... 199.9
200 ... 1999
0.1
1
ӕ
[mS/cm]
2.00 ... 19.99
20.0 ... 199.9
200 ... 500
0.01
0.1
1