In Programming mode, pressing the button decreases the displayed parameter value. 

B

ASIC 

O

PERATION

 

The Control Center Monitor has two modes: NORMAL RUN MODE and PROGRAMMING MODE.   

In Normal Run mode, the display shows the temperature that the probe is reading to the lowest 
whole number, i.e., if it shows a value of 10.0ᵒ, the actual temperature can range from 10.0 to 
10.9.   

Press the UP button to cause the display to show the high temperature set point.   

Press the DOWN button to cause the display to show the low temperature set point.   

The Control Center compares the probe’s temperature reading to HIGH and LOW temperature 
setpoints.    The  STATUS  LED  glows  green  when  the  probe  reads  an  in-range  temperature 
(between the two setpoints) and red when the temperature is out of range (equal to or outside 
the setpoints). 

The  Control  Center  will  not  declare  an  alarm  until  the  out-of-range  state  has  continued  for  a 
specified  time  (the  “alarm  delay”).    This  alarm  delay  time  is  designed  to  prevent  transient, 
nuisance alarms. 

When an alarm condition is declared, the BEEPER and ALARM LED will both activate.  Press the 
MUTE/PROGRAM button to stop the beeper for a specified time, changing the sound to short 
‘chirps’. When the temperature goes back within range, and there are no out-of-range readings 
for thirty seconds, the chirping stops and the MUTE/PROGRAM button resets.  If another out-
of-range condition then occurs, the beeper will again start at full volume.  

After the user-programmable “SILENCE” parameter times out, if the alarm condition still exists, 
the beeper will annunciate again at full volume until MUTE/PROGRAM is pushed. 

Battery Back-Up (B) 

The  Control  Center  has  a  battery  back-up  option.  It  ships  with  the  battery  in  place  and 
connected.    Remove  the  white  nylon  screw  located  on  the  lower  left  side  of  the  module  to 
activate the battery option.  

When  line  power  is  present,  all  of  the  normal  features  of  The  Control  Center  are  operating.  
When line power is lost, the battery provides power to maintain two important functions: 

 

Operate the audio and visual alarms.  They flash/beep less frequently than usual, in order to 
avoid draining the battery. 

 

Continue to display the temperature of the area being monitored.  The display will flash, to 
save power. 

 

 

Содержание FS-RFC-30

Страница 1: ...on the freight bill and acknowledged by the carrier s agent driver at time of delivery Make sure you get a signed copy Send a written request for an inspection to the carrier B Concealed Damage This...

Страница 2: ...NTY 22 BEFORE CALLING THE MANUFACTURER S TECHNICAL SUPPORT DEPARTMENT please have the unit s model and serial number ready as well as the problem description The model and serial number is located on...

Страница 3: ...Part number 20057 Rev 0 3...

Страница 4: ...y receipt acknowledging the damage This has to happen at the time of delivery or it won t happen at all Keep a copy for your records and send another to the carrier s damage claims department along wi...

Страница 5: ...tion VOLTAGE SHOULD NOT VARY MORE THAN 5 FROM SERIAL PLATE RATINGS A separate circuit is recommended to prevent possible loss of product due to over loading or failure of other equipment on the same c...

Страница 6: ...rly Temperature control NOTE Temperature controls are preset at the factory to provide the desired air and product temperatures inside of the units and require no further adjustment Please contact us...

Страница 7: ...top shelf avoid blocking the evaporator fan s if installed There should be at least 4 inches of clearance below the fan s to allow proper air flow Do not overload the unit A full load is about 70 to...

Страница 8: ...at the unit is in a power lock up mode and that the unit must be reset To reset the unit and clear the message from the display remove the plastic panel located below the solar panels by using a blunt...

Страница 9: ...ssociated with the alarm system i e delay times system status a battery backup which will maintain the alarm system for several hours after a power failure to allow for continued alarm and notificatio...

Страница 10: ...ible tick to indicate a button has been pressed The MUTE PROGRAM button has three functions In Normal Run mode if the unit is beeping pressing the button will mute the beeper for a preset time Each ti...

Страница 11: ...rms When an alarm condition is declared the BEEPER and ALARM LED will both activate Press the MUTE PROGRAM button to stop the beeper for a specified time changing the sound to short chirps When the te...

Страница 12: ...n a short period 5 Once the Monitor is in Programming mode each time the Mute button is pressed the display steps to the next parameter As the display steps to the next parameter the value of the prev...

Страница 13: ...m condition still exists the beeper will begin to sound at full volume again This parameter can be set from 5 to 120 minutes To view the Highest temperature HOLD both the Hi and the Lo buttons togethe...

Страница 14: ...s remote alarms or central monitoring systems that an alarm condition has occurred at the unit being monitored The relay will pull in when line power is present If an alarm condition lasts for a desig...

Страница 15: ...n is marking on the chart paper If it is not then carefully adjust the pen arm to establish contact with the paper REPLACEMENT OF PEN Note In non inking units that is when using pressure sensitive rec...

Страница 16: ...the time to adjust the position of the first pen has expired the second pen will move onto the chart briefly stopping at the outer most temperature graduation of the chart at which time the second pe...

Страница 17: ...he arrow button must be held down until the green LED light goes out After the green LED light goes out follow the instructions in step 3 above Refer to the following figure for a diagram of the push...

Страница 18: ...placements are pushed against rear wall or interrupted the proper refrigerator air circulation For the proper air circulation place the products evenly on each shelf Do not push against the refrigerat...

Страница 19: ...ent leakage Door will not close The unit is not level Refer to the Leveling section at the beginning of this manual Check for dirt and debris or items passing through the door seal MOISTURE DURING THE...

Страница 20: ...o not wash removable shelves in dishwasher Exterior and Handles Soap and water Non Abrasive Glass Cleaner Do not use commercial household cleaners ammonia or alcohol to clean handles Use a soft cloth...

Страница 21: ...ectly into a floor drain Replace the drain plug when defrosting is complete so warm air does not enter the freezer through the hole Defrosting tips If the frost is soft remove it with a plastic scrape...

Страница 22: ...Y In addition to the two 2 year warranty stated above Horizon Scientific Inc warrants its hermetically and semi hermetically sealed compressors to be free from defects in both material and workmanship...

Страница 23: ...od or other external causes It does not apply to defects resulting from failure to properly install operate or maintain the product in accordance with the printed instructions provided or damage cause...

Отзывы: