Application Hints
1.
Choose the cleaning method:
Direct cleaning:
place the support rack, water, detergent, and work piece into the
trough, or suspend the work piece in a bracket and submerge them into the
cleaning liquid. It is strictly prohibited to put alcohol, propylene, gasoline or other
flammable solvent, or strong acid or alkali, or other corrosive liquid into the trough.
If the above mentioned solvent is necessary then we recommend the use of indirect
cleaning method.
Indirect cleaning:
put water and detergent into trough, put all necessary chemicals
into the beaker or other proper container, and submerge them into the cleaning
liquid. Then put all the chemical detergent and work piece into the trough.
2.
The cleaner is powered by connecting the mains plug to a standard socket-outlet,
always place the cleaner in such a way that it is easy for the operator to unplug the
mains plug in emergency.
3.
Position your cleaner within easy reach of a standard grounded electrical outlet. Do
not place the cleaner on a circuit which could become overloaded.
4.
These cleaners are NOT floor standing equipment. Always use the cleaners on
tables, work benches, and other similar surfaces.
5.
Always use the cleaners in an environment of good ventilation. Special care should
be taken not to block the bottom fan vent of the relevant models.
6.
Please always keep the liquid level between the "
MAX.LEVEL
”and “
MIN.LEVEL
”.
7.
Adjust the timing to proper time. Work piece of different types require different
cleaning times. Most of them only need about a few minutes while others might
take a bit longer. The details can be decided via testing.
8.
It is faster and more efficient to run several small loads rather than few big loads.
9.
Turn power on,wait 10 minutes to let the cleaner liquid to purge the gas, the process
of which is only required at the beginning of every day cleaning or after the change
of solvent.
10.
When you first fill your unit, or refill it with fresh solution, use warm water for the
solution. Turn on the heater (press the HEAT switch, if available), turn on the
ultrasonics (press the SONICS), add the cover and the solution will heat quickly to
temperature.
11.
Rinse, drying and lubrication:
Rinse the parts to remove the chemicals which adhere to the parts after cleaning.
Dry the parts with clean compressed air, hot air blower or in an oven.
Relubricate parts that need lubrication.