Error and system messages | 9
B-H86.0.11.DB201-1.0
27 / 33
9 Error and system messages
Display
Meaning
Possible
causes
Remedy
>
(AL
<
Error during the last
calibration
Faulty calibration
Conduct a new calibration
No display,
unclear char-
acters or no
response
when but-
tons are
pressed
Battery depleted
System error
Product is defective
Battery depleted
Error in the product
Product is defective
Replace battery
Send in for repair
BAT
Battery depleted
Battery depleted
Replace battery
BAT LO
Battery depleted
Battery depleted
Replace battery
(AL ERR.1
Neutral buffer not al-
lowed
Incorrect buffer solu-
tion used
Buffer solution is
contaminated
Electrode contamin-
ated or defective
Use fresh buffer solution
Clean electrode, re-calibrate
Replace electrode
(AL ERR.2
Slope is too low
Incorrect buffer solu-
tion used
Buffer solution is
contaminated
Electrode contamin-
ated or defective
Use fresh buffer solution
Clean electrode, re-calibrate
Replace electrode
(AL ERR.3
Slope is too high
Incorrect buffer solu-
tion used
Buffer solution is
contaminated
Electrode contamin-
ated or defective
Use fresh buffer solution
Clean electrode, re-calibrate
Replace electrode
(AL ERR.4
Incorrect calibration
temperature
Temperature too low
or too high
Range of 0..60 °C
(AL ERR.5
Time exceeded dur-
ing automatic calib-
ration
Unstable electrode
signal
Buffer solution is
contaminated
Stirring of the buffer solution
Clean the electrode
Use fresh buffer solution
Restart calibration
ERR.1
Measuring range ex-
ceeded
Measurement too
high
Incorrect electrode
connected
Electrode or product
defect
The measurement is above
the permissible range
Check electrode
Send in for repair
ERR.2
Measuring range is
undercut
Measurement too
low
Incorrect electrode
connected
Electrode or product
defect
The measurement is below
the permissible range
Check electrode
Send in for repair
SYS ERR
System error
Error in the product
Switch product on/off