8
4.
Insert the final part of the projection lens into the
empty eyepiece sleeve. (Fig. 9)
5. In case of eyepiece sleeve with diameter 30 or
30.5 mm use the adapter rings. (Fig. 10)
8.2 Parfocality adjustment
In order to have the same focus when observing the
specimen through the eyepieces and on the screen/
monitor, please check that the microscope is properly
installed and follow the instructions below.
8.2.1 Using a biological microscope
1. Use a low power objective and focus the speci-
men.
2. Switch to the highest dry objective available on
the microscope (40x or 60x) and focus the speci-
men again.
3. Turn on the live-view on the camera, without
changing the focus on the microscope.
4. While observing the image on the screen/monitor,
adjust the focus by turning the knurled knob on
the C-mount adapter. (Fig. 11)
F
Fig. 9
ig. 9
F
Fig. 10
ig. 10
8.2.2 Using a stereomicroscope with “C” mount adapter
1.
Use a low power magnification and focus the
specimen.
2.
Reach the highest magnification available us
-
ing the zoom knob and then focus the specimen
again.
3. Turn on the live-view on the camera, without
changing the focus on the microscope.
4. While observing the image on the screen/monitor,
adjust the focus by turning the knurled knob on
the C-mount adapter. (Fig. 11)
F
Fig. 11
ig. 11