VWR
®
P Water System ver 2, rel 1 30.08.2019
Follow the same procedure to install other cartridges. Install the U Pack to the right side.
Note!
The control console will display
“RFID Error!”, and the alarm icon
if a pack is installed
incorrectly. Reinstall it to the correct position.
h) Make sure all consumables have been registered and tap
Next
to enter Dispenser setup
menu.
3.2.7 Dispenser Setup
Select connection mode for the dispenser from
the pull-down menu at the left corner:
Select
Wireless
for wireless connection.
Select
Wired
for the wired connection.
Tap
Query
, the system will automatically search
for the dispensers that have been installed.
Note:
The dispensers may have prior registered
to other systems, causing problems in installation
if record in the dispenser is not cleared. Press
Query will clear such record.
Tap
Reset
, system will automatically reset the
address.