VWR International 89508-420 Скачать руководство пользователя страница 11

Refrigerated Incubator

4-3

VWR Collection

Section 4

Unpacking and Installation

1.  Remove all protective tape from doors and shelves.

2.  Clean interior and exterior with warm soap and water. Shipping tape

residue can be removed using isopropyl alcohol.

3.  Uncoil line cord from rear of unit. Connect the plug to a power outlet

that complies with the electrical requirements specified on the unit's
label and with proper safety ground connection.

4.  Record serial number from unit on front cover of owner/operator

manual.

1.  Disconnect unit from power source. On inside of door, directly behind

controller, open snapped conduit and remove controller cable.

2.  Remove screw that secures cable clamp holding controller cable, then

replace screw.

3.  Remove controller cable from anchors and ties that run along inside

chamber side and back walls.

4.  Place 3 provided adhesive-backed mounting anchors at approximately

the same spacing (on opposite side) as existing anchors.

5.  While supporting door with one hand, remove screw securing door to

upper hinge, then lift door off bottom hinge.

6.  Remove the three screws securing each hinge to chassis.

7.  Remove three plug screws covering holes on left side, then install them

in holes on right side.

8.  Install hinges on left side using screws removed in Step 6. Position

hinge with door screw on bottom.

9.  Position door on hinge, then insert screw previously removed in Step

5.

10. Attach supplied latching duct to inside on door, in line with controller

cable toward opposite side on door from other duct.

11. Route cable using mounting anchors, ties and latching duct.

12. Remove screw on left front of enclosure and install cable clamp with

removed screw.

13. Adjust size of loop in cable where door is hinged to ensure door can

fully open and close without interference.

Installation
(continued)

Door Reversal

(if necessary)

Содержание 89508-420

Страница 1: ...User Manual VWR Collection Model 89508 420 Refrigerated 6 1 cu ft Incubator Operating and Maintenance Manual 7003734 Rev 0...

Страница 2: ...VWR Collection MANUAL NUMBER 7003734 0 10 7 13 Original ccs REV ECR ECN DATE DESCRIPTION By Preface Refrigerated Incubator i...

Страница 3: ...adjustments and maintenance must be performed by qualified service personnel s Material in this manual is for information purposes only The contents and the product it describes are subject to change...

Страница 4: ...E and the equipment that has been put on the market after 13 August 2005 This product is required to comply with the European Union s Waste Electrical Electronic Equipment WEEE Directive 2002 96 EC It...

Страница 5: ...andle Installation 4 4 Operation 5 1 General Functional Checks 5 2 Operation 5 2 Controller 5 2 Accessing Controller Menus 5 3 Changing Chamber Temp Control Setpoint 5 3 Maintenance 6 1 Cleaning 6 1 D...

Страница 6: ...ude Spacious 6 1 cu ft chamber accommodates up to 114 standard BOD bottles or similar containers Under the counter model with reversible door Two removable adjustable shelves 17 5 x 20 5 50 lbs weight...

Страница 7: ...t may be plugged into convenience outlet load current 2A at 115VAC leakage current 250 microamps 3 Disconnect unit power cord from the power supply prior to maintenance and servicing 4 When operating...

Страница 8: ...y 114 BOD bottles Shipping Weight 126 lb 57 kg Operating Environment Indoor Use Only Temperature Range 10 C to 40 C Humidity Range 0 to 90 RH Non Condensing Maximum Altitude 6600 ft 2 km Air Clearance...

Страница 9: ...the final carrier Usually the firm will send an investigator to ascertain liability Use the list below when unpacking to verify that the complete unit has been received Do not discard packing material...

Страница 10: ...quired leveler counterclockwise to increase the height and clockwise to reduce the height The Refrigerated Incubator should be installed on a level stable floor Caution Do not place unit on its top si...

Страница 11: ...back walls 4 Place 3 provided adhesive backed mounting anchors at approximately the same spacing on opposite side as existing anchors 5 While supporting door with one hand remove screw securing door...

Страница 12: ...While holding the handle in place tighten the screws with a 1 4 nut driver until they are tight against the top inner side of the door DO NOT OVERTIGHTEN SCREWS DO NOT GO INTO THE DOOR Section 4 Unpa...

Страница 13: ...located on front panel MENU Used to access menus SET Used to display setting s INCREASE setting t DECREASE setting Display on front panel Displays chamber and set point temperatures in C and prompts f...

Страница 14: ...ess SET to display control setpoint display should read 25 0 If the control setpoint or any of the other parameters aren t set correctly use the up or down arrow keys while pressing SET to change the...

Страница 15: ...lts arrange samples evenly throughout the chamber Liquid samples should be covered to prevent evaporation and eventual frost build up on evaporator coils particularly when operating below ambient Note...

Страница 16: ...ration that determines when the alarm is activated and if the control setpoint can be changed Range 00 to 02 00 Mode Normal mode that allows the user to change setpoint Alarm activates if chamber temp...

Страница 17: ...rature reaches the setpoint and is stable protection mode 02 can be set This mode can only be used if the door is to remain shut and other disturbances don t effect the chamber temperature As an added...

Страница 18: ...defrosts may be from 30 days to years Defrosting may be necessary when operating at 50 mode as well especially for control temperatures less than 15 0 C Method I For control setpoint down to 10 C thi...

Страница 19: ...an accurate temperature measuring device in the geometric center of the chamber 2 Set the controller to the desired operating temperature in 100 cooling mode and allow 2 hours for stabilization 3 Note...

Страница 20: ...lows Set DS1 1 A to off up Set DS1 2 B to off up 6 Attach wires previously removed from old controller to new controller 7 Mount new controller and bezel then apply power 8 Refer to Calibration sectio...

Страница 21: ...nect incubator from power source 2 Remove shelves from chamber area 3 Remove heater cover at top back of chamber by removing 2 screws on either side 4 Replace the THM1 or HTR1 if necessary then coat e...

Страница 22: ...VWR International LLC Radnor Corporate Center Building One Suite 200 100 Matsonford Road Radnor Pennsylvania 19087 United States 610 386 1700 www vwrsp com...

Отзывы: