10
light turns ON. Next turn the Safety clockwise just until the OTP turns OFF. Then
turn the Safety clockwise two (2) minor increments on its scale past the point where
the light went out. This sets the Overtemperature Safety Control at approximately
10
°
C above the Main Temperature set point. On “S” suffix models the reset button
will have to be pushed each time the Overtemperature Safety Control has been
activated and the OTP light comes on. On FM/GM models, the OTP light will come
on only when the heating element is activated, indicating that the temperature is
being maintained by the Overtemperature Safety.
6.6
Set Timer Display: Turn the timer switch to the ON position. The SET/TIMER
display digits will light with no lighted decimals showing (See Figure 2). Note that, if
during any of the following steps, several seconds elapse with no arrow pad or reset
pad activity, the timer will default to the present displayed setting and it will be
necessary to restart all functions over again. The values must be programmed in a
consecutive manner with no delays between settings or the default will occur.
A.
Hour
Function: Press and hold the RESET pad until the digits start blinking
and blinking decimal shows between digits 2 and 3. In this mode, pressing
the UP or DOWN arrow pads increases or decreases the whole hour value
from 0 to 99 (digits 1 and 2).
B.
Ten Minute Function: After the correct value for hours is set, push the
RESET pad again. The blinking decimal will now move one digit to the right
between digits 3 and 4. Pushing the UP or DOWN arrow pads will increase
or decrease the ten minute function allowing values between 0 and 5 to be
set (digit 3).
C.
One
Minute
Function: After the correct ten minute value is set, push the
RESET pad again. The blinking decimal point will move one digit to the right
beyond digit 4 and be located at the extreme bottom right of the display.
With the display in this mode, pushing the UP or DOWN arrow pad will
increase or decrease the one minute function allowing the value of digit 4 to
be adjusted between 0 and 9.
D.
Activation: Pause until the timer stops blinking. After all settings are made,
push the START/STOP button. The Timer Activated light will come on and
after a brief pause, the present oven temperature settings will be valid and
heating will begin. The oven will now heat up, control at the set point and
stop after the timed period on the SET/TIMER display has elapsed.
Note that when the system is in the timer mode, the heating circuit is de-
energized until the START/STOP button is pushed or the TIMER SWITCH is
turned O/Off. If a time change or correction is necessary and the timer has
already been activated, push the START/STOP button to “STOP” the timer,
then repeat steps A through D above.
6.7
To set the timer so that timed operation will not start until the oven is stable at set
point, pre-heat the oven in the normal mode until the desired temperature has
Содержание 1330FM