5.6 DIAGNOSTIC TABLES
ANALOG CIRCUITRY
Fault
Setup
Check
Results desired
Muffle heats but
N/A
YELL2
TC yellow connected
display shows
U4
Gain approx. 20
same temperature
D1
1.23V
Erratic temperature
D1
1.23V stable
display
N/A
U4-11 & 12
DC stable
Temperature drift
Hi T = 960°C
RED
39-40mV stable
U4-13
mV stable
(.2mV/°C ambient
increase typical)
Turn furnace off
Reconnect
No Err2 at elevated
TC wires
temperatures
DISPLAY BOARD
Fault
Setup
Check
Results desired
LCD dots are all
Furnace in idle
J8-1
+5V
on or off at power up
J4-8
Tight fit
No LED's light up
Perform power up
Front Panel
LED's turn off one
by one
N/A
J8-4
+12V
One LED does not come on
Turn furnace off
LED
Not open or
shorted
LCD display is darker
Turn furnace off
Temperature on
Less than 40°C
panel
Turn furnace on
Temperature on
LCD lighter shade
panel
SECTION 5 - TROUBLESHOOTING - 3-STAGE