VTI Instruments Corp.
54
EX1629 Basic Operation
The default settings are: Butterworth filter with -3 dB freq = 10 Hz, bilinear transform, and user
order = 0 (calculated order = 6). More information on filtering can be found in
Appendix B
T
RIGGERING
The EX1629 supports a full function trigger model with a separate arm source and trigger source
event structure. For a complete explanation of the trigger model, see Section 4. In summary, an
acquisition sequence is enabled with a trigger initialize function. Measurement data is then
acquired upon the receipt of the programmed arm source event followed by the receipt of the
programmed trigger source event. Trigger and arm source events can be independently
programmed from a variety of sources including Immediate, Timer, Digital I/O, and the Trigger
Bus.
D
ATA
F
ORMAT
By default, the data returned during data retrieval is limited to the main bridge readings and the
absolute time of scan initiation. If enabled, the EX1629 can also return data from its confidence
measurement system, such as the excitation source voltages and excitation source currents.
S
HUNT
C
ALIBRATION
The EX1629 features an extremely capable and fully programmable shunt calibration architecture
to ensure correct bridge performance. Three discrete shunt calibration modes are supported that
can be employed in local and remote connections. Figure 3-8 illustrates the shunt calibration
design.
TEDS
COMM
External customer
shunt cal
resistor terminals
Internal
Rcal
55K
-R Cal
+R Cal
Pin 5
Pin 4
Strain
Bridge
8-pin telcom
connector (RJ-45)
350
120
User
10K
10K
+Excitation Sense
+ Excitation
- Excitation
Full-Bridge
(1 of 48 channels)
+Excitation
+Sense
-Sense
+Excitation Sense
-Excitation Sense
-Excitation
-Excitation Sense
RJ-45 Pin#
12345678
Pin 2
Pin 7
Pin 6
Pin 3
Pin 8
Pin 1
+
-
-
+
D
C
B
E
F
A
10K
10K
Instrumentation
Amplifier
+
-
x1
x10
x100
Hi
Lo
Shield
150
150
-Sense
-Cal
AGND
+Sense
Wagner
+Cal
AGND
430 kHz
LOW-PASS
FILTER
60.3 kHz
LOW-PASS
FILTER
ADC
TRIFILAR
FILTER
F
IGURE
3-8:
S
HUNT
C
ALIBRATION
C
ONFIGURATION