www.vtiinstruments.com
EX1401 Calibration
107
O
BTAIN
2.5V
REFERENCE VOLTAGES
This process is for the verification of reference voltages used within the EX1401 instrument.
7.
In the Soft Front Panel, open ‘Internal Calibration References’ window. Enable all channels
by individually selecting the ‘Enabled’ checkboxes as shown in
Error! Reference source not
found.
.
F
IGURE
8-1:
I
NTERNAL
C
ALIBRATION
R
EFERENCES
W
INDOW
8.
In the ‘Channels’ window, enable all channels and configure them to ‘Cal’ function, 10V
range and 10000 Record Size as shown in
Error! Reference source not found.
.
F
IGURE
8-2:
I
NTERNAL
C
AL
R
EFERENCES
:
C
HANNELS
W
INDOW
9.
Ensure that Clock and Sample Rate are set to 10000 in the ‘Sampling’ window.
10.
Click ‘Initiate Measurement’ button in ‘Measurement’ window and wait for the acquisition to
complete. With Download Records set to 1, click ‘Download CSV’.
11.
Open the downloaded csv file. For each channel, average the 10,000 samples and write down
the values.
12.
Return all functions back to Voltage in the ‘Channels’ window and disable all channels in the
‘Internal Calibration References’ window.
Содержание EX1401
Страница 2: ......
Страница 24: ......
Страница 57: ...www vtiinstruments com EX1401 Web Page Operation 57 ...
Страница 94: ......
Страница 98: ......
Страница 102: ......
Страница 111: ...www vtiinstruments com EX10xxA Index 111 FIGURE A 1 IEEE1588 NETWORK TOPOLOGY ...
Страница 112: ......
Страница 114: ...VTI Instruments Corp 114 EX1401 Index ...