VTI Instruments Corp.
18
EX1200 Series Terminal Blocks
2)
Save and provide a name for the DMM configuration for the scanning purposes (here,
TEMP
is used).
3)
Setup the scanlist
a. Add a scan step with a sweep phase that scans CH1_1 using the previously saved
configuration (e.g. CH1_1=
TEMP
)
b. The setup phase can be ignored for this example.
4)
Initiate & Read
a. Initiate the scan
b. Read back the data
Use Thermistor as a Reference Junction
1)
Set the DMM to measure temperature using the thermistor.
a. Set the DMM function to
Temperature
.
b. Set the transducer type to
Thermistor
.
c. Set the thermistor resistance to
10000
(the thermistor is 10 kΩ).
d. Set the
isReferenceJunction
parameter to
True
.
2)
Save and provide a name for the DMM configuration for the scanning purposes.
3)
Set the DMM to measure temperature using a Type K thermocouple (any thermocouple type
can be selected here, but Type K below).
a. Set the DMM function to
Temperature
.
b. Set the transducer type to
Thermocouple
.
c. Set the thermocouple type to
Type K
.
d. Set the reference type to
External
.
e. Set the
isReferenceJunction
parameter to
False
.
4)
Save and provide a name for the DMM configuration for the scanning purposes.
5)
Setup the scanlist
a. Add a scan step with a sweep phase that scans CH1_1 using the first saved configuration.
(e.g. CH1_1=
RefJunction
).
b. Add another scan step with a sweep phase that scans another channel (e.g.
CH1_2=
TEMP
).
c. The setup phase can be ignored for this example.
6)
Initiate & Read
a. Initiate the scan
b. Read back the data
Содержание EX1200 Series
Страница 21: ...www vtiinstruments com EX1200 2001 Plug in Module 21 LOGICAL DIAGRAM FIGURE 4 2 EX1200 2001 LOGICAL DIAGRAM ...
Страница 32: ...VTI Instruments Corp 32 EX1200 2008 HPlug in Module LOGICAL DIAGRAM FIGURE 4 9 EX1200 2008H LOGICAL DIAGRAM ...
Страница 36: ...VTI Instruments Corp 36 EX1200 3001 Plug in Module LOGICAL DIAGRAM FIGURE 4 11 EX1200 3001 LOGICAL DIAGRAM ...
Страница 45: ...www vtiinstruments com EX1200 3048 Plug in Module 45 LOGICAL DIAGRAM FIGURE 4 15 EX1200 3048 LOGICAL DIAGRAM ...
Страница 61: ...www vtiinstruments com EX1200 3096 Plug in Module 61 LOGICAL DIAGRAM FIGURE 4 21 EX1200 3096 LOGICAL DIAGRAM ...
Страница 66: ...VTI Instruments Corp 66 EX1200 3164 Plug in Module LOGICAL DIAGRAM FIGURE 4 23 EX1200 3164 LOGICAL DIAGRAM ...
Страница 79: ...www vtiinstruments com EX1200 4003 Plug in Module 79 LOGICAL DIAGRAM FIGURE 4 32 EX1200 4003 LOGICAL DIAGRAM ...
Страница 99: ...www vtiinstruments com EX1200 4264 Plug in Module 99 LOGICAL DIAGRAM FIGURE 4 39 EX1200 4264 LOGICAL DIAGRAM ...
Страница 103: ...www vtiinstruments com EX1200 5001 Plug in Module 103 LOGICAL DIAGRAM FIGURE 4 42 EX1200 5001 LOGICAL DIAGRAM ...
Страница 113: ...www vtiinstruments com EX1200 5004 Plug in Module 113 LOGICAL DIAGRAM FIGURE 4 46 EX1200 5004 LOGICAL DIAGRAM ...
Страница 117: ...www vtiinstruments com EX1200 5006 Plug in Module 117 LOGICAL DIAGRAM FIGURE 4 48 EX1200 5006 LOGICAL DIAGRAM ...
Страница 122: ...VTI Instruments Corp 122 EX1200 6101 Plug in Module LOGICAL DIAGRAM FIGURE 4 50 EX1200 6101 LOGICAL DIAGRAM ...
Страница 126: ...VTI Instruments Corp 126 EX1200 6102 Plug in Module LOGICAL DIAGRAM FIGURE 4 52 EX1200 6102 LOGICAL DIAGRAM ...
Страница 130: ...VTI Instruments Corp 130 EX1200 6111 Plug in Module LOGICAL DIAGRAM FIGURE 4 54 EX1200 6111 LOGICAL DIAGRAM ...
Страница 138: ...VTI Instruments Corp 138 EX1200 6216HV Plug in Module LOGICAL DIAGRAM FIGURE 4 58 EX1200 6216HV LOGICAL DIAGRAM ...
Страница 141: ...www vtiinstruments com EX1200 6301 Plug in Module 141 LOGICAL DIAGRAM FIGURE 4 60 EX1200 6301 LOGICAL DIAGRAM ...
Страница 144: ...VTI Instruments Corp 144 EX1200 6301T Plug in Module LOGICAL DIAGRAM FIGURE 4 62 EX1200 6301T LOGICAL DIAGRAM ...
Страница 166: ......