EX10xxA Triggering
59
The measurement range of the EX10xxA in terms of temperature is a function of its input voltage
range and the capabilities of the thermocouple sensors themselves. Specifically, the measurement
range of the EX10xxA for the standard thermocouple types is the following:
Type
Min (°C)
Max (°C)
Min (°F)
Max (°F)
J
-200
1200
-328
2192
K
-200
1372
-328
2502
T
-200
400
-328
752
E
-200
900
-328
1652
S
-50
1768
-58
3214
R
-50
1768
-58
3214
B
250
1820
482
3308
N
-200
1300
-328
2372
T
ABLE
3-2:
EX10
XX
A
M
EASUREMENT
R
ANGE
In addition to the standard thermocouple types, the EX10xxA can accept and measure a custom
thermocouple of any type. The effective measurement range in temperature is subsequently
determined by the input voltage range of the EX10xxA and the thermocouple transfer function of
the custom thermocouple.
In terms of voltage, the thermocouple connectors may only accept an input voltage of ±67 mV when
performing temperature conversion. When in voltage mode, these connectors can accept either
±0.067 V or ±0.1 V inputs. The voltage connectors may use either ±0.01 V, ±0.067 V, ±0.1 V,
±1.0 V, or ±10.0 V inputs for both EU temperature conversion or voltage measurements. When
temperature is selected, the instrument automatically switches the selected channel(s) to the ±67 mV
range. The EX1000A-TC is a special case where thermocouple inputs are used for voltage
measurement and may utilize all input ranges available to the voltage connectors. Table 1- provides
an overview of each models capabilities.
C
OLD
J
UNCTION
C
OMPENSATION
(CJC)
For highest accuracy and stability, the EX1000A-TC, EX1016A, EX1032A, EX1048A and RX1032
provide embedded isothermal input sections that are monitored by twelve precision thermistors
(based on Model#, number of thermistors may vary), one for every four thermocouple channels.
(Note, the EX1000A may utilize external thermistors as described in the
section). To ensure that CJC information is current and time correlated with the input channels, CJC
channels are measured with every scan, providing a maximum time separation of less than 4 ms
between the input channel measurement and its associated CJC measurement. Table 3-3 shows the
association between the CJC and input channels.
CJC #
Channel
Input Channel
CJC0
48
0-3
CJC1
49
4-7
CJC2
50
8-11
CJC3
51
12-15
CJC4
52
16-19
CJC5
53
20-23
CJC6
54
24-27
CJC7
55
28-31
CJC8
56
32-35
CJC9
57
36-39
CJC10
58
40-43
CJC11
59
44-47
T
ABLE
3-3:
CJC
C
HANNEL
/
I
NPUT
C
HANNEL
R
ELATIONSHIP
Содержание EX1000A
Страница 25: ...www vtiinstruments com EX10xxA Introduction 25 FIGURE 1 2 RACKMOUNT HARDWARE INSTALLATION DIAGRAM ...
Страница 27: ...www vtiinstruments com EX10xxA Introduction 27 FIGURE 1 4 ISOMETRIC AND INTERNAL CJC PLATE DIAGRAM ...
Страница 28: ...VTI Instruments Corp 28 EX10xxA Introduction EX1044 DIAGRAM ...
Страница 29: ...www vtiinstruments com EX10xxA Introduction 29 FIGURE 1 5 EX1044 TABLE TOP USAGE ...
Страница 34: ......
Страница 56: ......
Страница 74: ......
Страница 130: ......
Страница 282: ......