Performance Characteristics
53
Regression statistics
Within run precision
(100 venous blood tests per glucose level)
Total precision
(100 control solution tests per glucose level)
Slope
0.9995
Intercept
1.0856 mg/dL
R
2
0.9874
No. of samples
100
Range tested
33.6~433 mg/dL (1.8~24.0 mmol/L)
Mean glucose
Standard
deviation
Coefficient
of variation
(%)
41 mg/dL (2.3 mmol/L)
1.3 mg/dL
(0.07 mmol/L)
3.1
99 mg/dL (5.5 mmol/L)
2.8 mg/dL
(0.16 mmol/L)
2.9
131 mg/dL (7.3 mmol/L)
2.9 mg/dL
(0.16 mmol/L)
2.2
231 mg/dL (12.8 mmol/L)
4.3 mg/dL
(0.24 mmol/L)
1.9
373 mg/dL (20.7 mmol/L)
9.8 mg/dL
(0.54 mmol/L)
2.6
Mean glucose
Standard
deviation
Coefficient
of variation
(%)
43 mg/dL (2.3 mmol/L)
1.1 mg/dL
(0.06 mmol/L)
2.5
104 mg/dL (5.7 mmol/L)
2.4 mg/dL
(0.13 mmol/L)
2.3
299 mg/dL (16.6 mmol/L)
8.3 mg/dL
(0.46 mmol/L)
2.8
Содержание 2 in 1 Smart
Страница 29: ...Testing Your Blood Glucose 29 STEP 5 Touch the drop to test strip tip Blood sample is drawn into test strip...
Страница 30: ...Testing Your Blood Glucose 30 The volume of blood sample must be at least 0 5 microliter real size IMPORTANT...
Страница 88: ...88 Testovanie hladiny gluk zy v krvi Objem krvnej vzorky mus by minim lne 0 5 mikrolitra skuto n ve kos D LE IT...