QUICK START GUIDE - USING MULTIPLE DC1000S WITH A LCR METER 

 

PAGE 3 OF 3 

104-174/1 

 

 

D Running 3 units in parallel 

Once you have compensated the LCR for the leads and DC1000s you 
can now make measurements under load. 

 

1 Turn the DC1000 output 

OFF. 

2 Add your 

UUT

 back into the circuit, attempting to keep the test leads 

in the 

same position

 as when compensating. 

3 Select the 

DC Bias Current

 you require using the rotary knob on the 

MASTER

 unit. The Slave units will adjust automatically. In the example 

shown each unit is providing 

15.00 AMPS

 to give a total DC bias 

current of 

45.00 AMPS.

 

4 Enable the DC Bias Current by pressing 

OUTPUT

 on the 

MASTER

 

unit 

5 Perform measurements on the 

LCR.

 

6 Adjustments can be made to the DC BIAS level while the output is 
enabled (and the LCT is measuring) using the 

rotary knob

.

 

 

 

E Best Practice / Warnings

 

 

1

 DO NOT

 disconnect the UUT or LCR whilst the DC1000 is operating. 

2 Always remove the DC BIAS current by 

disabling the OUTPUT

 button before disconnecting 

either the UUT or the LCR meter. 

3 Always try to keep all leads in the 

same position

 to improve the accuracy of the 

compensation, and hence the accuracy of your measurements. 

 

 