C H A P T E R 4
–
A T 5 6 0 0 F U N C T I O N A L D E S C R I P T I O N
AT5600 User Manual 98-119 issue 14
Page 57
Low Power Generator
–
50 Ohm impedance
The Low Power Generator can produce up to 5 V
RMS
and contains
50Ω source
impedance, therefore the maximum Test Signal which can be delivered depends
on the load connected to the source. The maximum Test Signal shall be;
V
out
= V
LPG
* [Z
Load
/ (Z
Load
+ R
50Ω
)]
Example:
An
“
R Test
”
to measure a component with a nominal resistance of 500
Ω, the
maximum Test Signal that can be programmed is;
V
out
= 5V * [500
/ (500 + 50)] = 4.55 V
4.1.5.
Measurement Circuits
The AT5600 tester contains a variety of circuits, capable of performing all the
following measurements:
•
Voltages from less than 1mV to greater than 7000V
•
Currents from nano-amps to amps
•
DC and harmonic analysis
•
RMS, mean-sense, wattage and VA analysis
Again, the processors select which is appropriate for each test as it is executed.
Содержание AT5600
Страница 10: ...AT5600 User Manual 98 119 issue 14 Page 10 16 4 AT Series Server Software 437...
Страница 39: ...C H A P T E R 2 S Y S T E M O V E R V I E W AT5600 User Manual 98 119 issue 14 Page 39...
Страница 88: ...C H A P T E R 6 S A F E T Y S Y S T E M S AT5600 User Manual 98 119 issue 14 Page 88...
Страница 168: ...C H A P T E R 7 T E S T S A N D T E S T C O N D I T I O N S AT5600 User Manual 98 119 issue 14 Page 168...
Страница 198: ...C H A P T E R 8 F R O N T P A N E L O P E R A T I O N AT5600 User Manual 98 119 issue 14 Page 198...
Страница 204: ...C H A P T E R 9 T R O U B L E S H O O T I N G AT5600 User Manual 98 119 issue 14 Page 204...
Страница 252: ...C H A P T E R 1 0 S P E C I F I C A T I O N S AT5600 User Manual 98 119 issue 14 Page 252...
Страница 258: ...C H A P T E R 1 2 W A R R A N T Y A N D S E R V I C E AT5600 User Manual 98 119 issue 14 Page 258...
Страница 270: ...C H A P T E R 1 3 T E S T F I X T U R E S AT5600 User Manual 98 119 issue 14 Page 270...