C H A P T E R 1 6 - C H A N G E L O G
AT5600 User Manual 98-119 issue 14
Page 428
16.2.
AT5600 Firmware
V 1.001.026
–
18 FEB 2019
FUNCTIONALITY
COMMS
When printing failures only to USB printer, the AT now prints the test number not the
row position. (ID 9060)
This version also takes care of the editor locking up when running a program on the
AT5600 while the safety interlock is already broken
Improved the TCP/IP connection process to be more robust.
Safety aborts are no longer latched while the Kelvin failure screen is shown.
The RUN led on the front panel is no longer left illuminated while the printing and
communications are in progress.
GENERAL
Stop on fail is now followed by a safety interlock check.
Kelvin checks are now followed by a safety interlock check.
Resolved an issue that caused the AT5600 to crash if the printer ran out of paper.
The remote BUSY light now turns on appropriately.
Changed the text 'Transformers Tested' to 'Programs Run' on the program execution
screen.
Stop on fail now uses the user set failure beep setting.
Added specification-based integration for CTY, R, R2, LS, LP, L2, QL, D, RLS, RLP,
Z, ANGL, TR and LVOC to improve speed
Fixed an issue with self-test ID 3100 false failing
Added a method to detect and recover from disruption on the USB peripheral.
Added the ability to move to a non-ideal resistance range combination when in high
noise environments.
TESTS
HPAC
Modified the trip level for the AC-EHT generator to allow it to execute
high voltage high frequency tests without tripping.
HPAC HPDC Improved the trimming for low level class D, HPDC and HPAC tests to
reduce the chances of a spurious 0020 error code.
LL
Deleted a legacy correction coefficient on the leakage inductance test.
LSB LPB ZB Compensation and Dc bias trimming is now improved for LPB and ZB
tests.
LSBX LPBX ZBX Compensation is now improved or LSBX, LPBX and ZBX tests.
LS LP
The accuracy has been improved at 1MHz and up. This improvement
also requires re-calibration
LVOC
LVOC test signal trim improvement
LVOC
LVOC tests now use turns ratios compensation method.
MAGI
Mean sense RMS scaled in MAGI now works properly.
MAGI VOC WATT The ranging and learning process for class D tests has been
modified to prevent false overloads.
MAGX
MAGX tests now re-use energization properly if the energized test is
the first test.
R CTY
Improved the ranging process for DCR to prevent range hunting.
Содержание AT5600
Страница 10: ...AT5600 User Manual 98 119 issue 14 Page 10 16 4 AT Series Server Software 437...
Страница 39: ...C H A P T E R 2 S Y S T E M O V E R V I E W AT5600 User Manual 98 119 issue 14 Page 39...
Страница 88: ...C H A P T E R 6 S A F E T Y S Y S T E M S AT5600 User Manual 98 119 issue 14 Page 88...
Страница 168: ...C H A P T E R 7 T E S T S A N D T E S T C O N D I T I O N S AT5600 User Manual 98 119 issue 14 Page 168...
Страница 198: ...C H A P T E R 8 F R O N T P A N E L O P E R A T I O N AT5600 User Manual 98 119 issue 14 Page 198...
Страница 204: ...C H A P T E R 9 T R O U B L E S H O O T I N G AT5600 User Manual 98 119 issue 14 Page 204...
Страница 252: ...C H A P T E R 1 0 S P E C I F I C A T I O N S AT5600 User Manual 98 119 issue 14 Page 252...
Страница 258: ...C H A P T E R 1 2 W A R R A N T Y A N D S E R V I C E AT5600 User Manual 98 119 issue 14 Page 258...
Страница 270: ...C H A P T E R 1 3 T E S T F I X T U R E S AT5600 User Manual 98 119 issue 14 Page 270...