46
12. Overview of operation modes, settings and
LC Display indications
a) Operation modes and settings
For performing following configuration settings, please refer to “MEASUREMENT” options - “BASIC SETTING” on
page 41.
Mode 1
Function
Logging starts immediately.
Logging stops when memory is full.
Setting
Select “IMMEDIATELY UNTIL MEMORY FULL” as “RECORD START CONDITION”.
Mode 2
Function
Logging starts immediately.
Logging stops when button ENTER (7) is pressed for few seconds.
Setting
Select “IMMEDIATELY UNTIL MEMORY FULL” as “RECORD START CONDITION”.
Select “STOP BUTTON” as “STOP STYLE”.
Mode 3
Function
Logging starts immediately.
Logging stops after PDF report generation.
Setting
Select “IMMEDIATELY UNTIL MEMORY FULL” as “RECORD START CONDITION”.
Select “AFTER PDF CREATED” as “STOP STYLE”.
Mode 4
Function
Logging starts when button ENTER (7) is pressed for few seconds.
Logging stops when memory is full.
Setting
Select “START UPON KEYPRESS” as “RECORD START CONDITION”.
Mode 5
Function
Logging starts and stops when button ENTER (7) is pressed for few seconds.
Setting
Select “START UPON KEYPRESS” as “RECORD START CONDITION”.
Select “STOP BUTTON” as “STOP STYLE”.
Mode 6
Function
Logging starts when button ENTER (7) is pressed for few seconds.
Logging stops after PDF report generation.
Setting
Select “START UPON KEYPRESS” as “RECORD START CONDITION”.
Select “AFTER PDF CREATED” as “STOP STYLE”.