41
COMPARING METER AND LABORATORY
RESULTS
Before you go to the lab:
Perform a control solution test to make sure that the
▪
meter is working properly.
Fast for at least eight hours before doing comparison
▪
tests, if possible.
Take your meter with you to the lab.
▪
Make sure that the samples for both tests are taken and
tested within 15 minutes of each other.
Wash your hands before obtaining a blood sample.
▪
Never use your meter with blood that has been
▪
collected in a gray-top test tube.
Use fresh capillary blood only.
▪
Test results from the meter and laboratory are both
expressed in plasma-equivalent units. The result you
obtain from your meter may differ somewhat from your
laboratory result due to normal variation. Meter results
can be affected by factors and conditions that do not
affect laboratory results in the same way (See test strip
package insert for typical accuracy and precision data,
and for important information on Limitations). To make
an accurate comparison between meter and laboratory
results, follow the guidelines below.