QA Maintenance and Inspection ViZion DR+ Detector Guide  v2.0 

 

Opal

 

4

 

 
 

 
 

 

 

 

 

Customer Support 

1.800.366.5343 – 

[email protected]

 

VZDR-UG-QA Maintenance and Inspection ViZion DR+ Detector Guide-D-20141012-2.0 

VIII. 

Monthly Inspection 

Perform the following inspection periodically at least once a month. Contact a VIZION DR+ Dealer 
representative or distributor if there is any problem. 

 

Phantom Inspection:  

1.  Visually check the performance of the instrument by performing 

exposures using a phantom or a resolution chart in 4 corners of the 
image and in the center of the image. Figure 1 

 

2.  Record the maximal distinguishable resolution.  

 

3.  If maximal distinguishable resolution is less than 3.4, repeat the 

calibration. 

 
 

Data 

Middle 

top of left 

top of right 

bottom of left 

bottom of right 

Installation 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Flat field Inspection:  
 
Check the performance of the instrument by performing the following 

Flat Field Inspection of ViZion 

DR+  Instructions 

and document the Date and Results. 

Date: 

Results: 

 

 

Figure 1

 